User Benefits
Application
News
◆ Validated method for 30 PFAS in Fish meeting all criteria of AOAC SMPR 2023.003
◆ High precision, excellent recovery, low Limit of Quantification (LOQ)
◆ Simple and rapid extraction using QuEChERS
LCMS -8060NX High Performance Liquid Chromatograph Mass Spectrometer
Nexera series High Performance Liquid Chromatograph
Determination of 30 PFAS in Fish by Liquid
Chromatography Triple Quadrupole Mass
Spectrometry (LC-MS/MS)
■ Introduction
Per-and polyfluorinated substances (PFAS) are a diverse
group of man-made chemicals used in numerous products
since the 1950s. PFAS can enter the food supply by
contact in environmentally contaminated areas, during
food processing, or exposure to packaging. Because PFAS
have been linked to serious health effects, accurate
methodology is needed. In this application news, we
describe a single laboratory validation study with a rapid
extraction of low concentrations of 30 PFAS in fish using
the QuEChERS technique followed by analysis using the
Shimadzu Nexera Liquid Chromatograph coupled to a
Shimadzu LCMS-8060NX triple quadrupole mass
spectrometer (Figure 1).
We optimized the chromatography and instrument
operating parameters to achieve excellent peak shape,
separation, and sensitivity. Sensitivity was improved for 4
PFAS, PFOA, PFHxS, PFNA, and PFOS.
In this study, we spiked samples at three concentrations in
triplicate. For greater accuracy, standards were matrix-
matched and extracted and spikes were quantified using
the isotope dilution technique. Recovery and precision
were compared to the requirements of AOAC SMPR
2023.003. In addition, we determined the Limit of
Quantitation (LOQ) as the lowest concentration meeting
accuracy and precision, ion ratio, retention time, and
signal-to-noise ratio criteria of the qualifier ion. All
recovery, precision, and LOQ’s met the acceptance criteria
of the SMPR. The target analytes, their acronym, chemical
abstract number, and experimentally determined LOQ are
shown in Table 1.
Figure 1: Nexera and LCMS -8060NX. The ion focus design
improves signal intensity with higher gas flows and higher effective
temperatures.
Analyte name Acronym CAS No. LOQ (ppb)
Perfluorobutanoic acid PFBA 375-22-4 0.1
Perfluoropentanoic acid PFPeA 2706-90-3 0.1
Perfluorohexanoic acid PFHxA 307-24-4 0.1
Perfluoroheptanoic acid PFHpA 375-85-9 0.1
Perfluorooctanoic acid PFOA 335-67-1 0.1
Perfluorononanoic acid PFNA 375-95-1 0.1
Perfluorodecanoic acid PFDA 335-76-2 0.1
Perfluoroundecanoic acid PFUnA 2058-94-8 0.1
Perfluorododecanoic acid PFDoA 307-55-1 0.1
Table 1: PFAS Analytes, Acronyms, CAS No. and Method LOQ
1
Shimadzu Scientific Instruments, Inc.,
2
Shimadzu Corporation
William Lipps
1
, Toshiya Matsubara
1
, Dominika Gruszecka
1
, Nozomi Maeshima
2
, Kota Ishioka
2
,
Manami Kobayashi
2