User Benefits Application News ◆ Validated method for 30 PFAS in Fish meeting all criteria of AOAC SMPR 2023.003 ◆ High precision, excellent recovery, low Limit of Quantification (LOQ) ◆ Simple and rapid extraction using QuEChERS LCMS -8060NX High Performance Liquid Chromatograph Mass Spectrometer Nexera series High Performance Liquid Chromatograph Determination of 30 PFAS in Fish by Liquid Chromatography Triple Quadrupole Mass Spectrometry (LC-MS/MS) ■ Introduction Per-and polyfluorinated substances (PFAS) are a diverse group of man-made chemicals used in numerous products since the 1950s. PFAS can enter the food supply by contact in environmentally contaminated areas, during food processing, or exposure to packaging. Because PFAS have been linked to serious health effects, accurate methodology is needed. In this application news, we describe a single laboratory validation study with a rapid extraction of low concentrations of 30 PFAS in fish using the QuEChERS technique followed by analysis using the Shimadzu Nexera Liquid Chromatograph coupled to a Shimadzu LCMS-8060NX triple quadrupole mass spectrometer (Figure 1). We optimized the chromatography and instrument operating parameters to achieve excellent peak shape, separation, and sensitivity. Sensitivity was improved for 4 PFAS, PFOA, PFHxS, PFNA, and PFOS. In this study, we spiked samples at three concentrations in triplicate. For greater accuracy, standards were matrix- matched and extracted and spikes were quantified using the isotope dilution technique. Recovery and precision were compared to the requirements of AOAC SMPR 2023.003. In addition, we determined the Limit of Quantitation (LOQ) as the lowest concentration meeting accuracy and precision, ion ratio, retention time, and signal-to-noise ratio criteria of the qualifier ion. All recovery, precision, and LOQ’s met the acceptance criteria of the SMPR. The target analytes, their acronym, chemical abstract number, and experimentally determined LOQ are shown in Table 1. Figure 1: Nexera and LCMS -8060NX. The ion focus design improves signal intensity with higher gas flows and higher effective temperatures. Analyte name Acronym CAS No. LOQ (ppb) Perfluorobutanoic acid PFBA 375-22-4 0.1 Perfluoropentanoic acid PFPeA 2706-90-3 0.1 Perfluorohexanoic acid PFHxA 307-24-4 0.1 Perfluoroheptanoic acid PFHpA 375-85-9 0.1 Perfluorooctanoic acid PFOA 335-67-1 0.1 Perfluorononanoic acid PFNA 375-95-1 0.1 Perfluorodecanoic acid PFDA 335-76-2 0.1 Perfluoroundecanoic acid PFUnA 2058-94-8 0.1 Perfluorododecanoic acid PFDoA 307-55-1 0.1 Table 1: PFAS Analytes, Acronyms, CAS No. and Method LOQ 1 Shimadzu Scientific Instruments, Inc., 2 Shimadzu Corporation William Lipps 1 , Toshiya Matsubara 1 , Dominika Gruszecka 1 , Nozomi Maeshima 2 , Kota Ishioka 2 , Manami Kobayashi 2
■ Sample Preparation and Analysis Conditions Samples were prepared by dicing the edible portion of tuna filets and grinding with dry ice. The ground material was placed in a freezer overnight. Test portions were spiked in triplicate at three different concentrations with 30 native PFAS (Table 1) and 16 isotopically labeled internal standards. Calibration curves for use in the quantitative analysis were prepared using 10- gram test portions spiked with concentrations of 0.05, 0.1, 0.5, 1.0 and 5.0 ng/g. Quantitation was carried out on additional tuna samples spiked in triplicate at 0.1, 1.0, and 5.0 ng/g. Since standards were extracted in a tuna matrix and carried through the same procedure, the final concentration of each PFAS in the sample can be calculated directly from the curve. 10-gram portions were weighed, spiked with target analytes and internal standards, and 10 mL of acetonitrile was added. The samples were shaken by hand for 1 minute and a QuEChERS packet was added. The sample was shaken for 1 minute and then centrifuged for 5 minutes at 4000 rpm. An aliquot of the acetonitrile layer was transferred to a tube and diluted 5 times with PFAS-free reagent water. The sample was then passed through a weak anion exchange (WAX) Solid Phase Extraction (SPE) cartridge and the PFAS were eluted with basic methanol. A 0.5 mL aliquot was transferred to a LC vial and acidified with formic acid. A volume suitable to obtain the required sensitivity of the extract was injected onto a UHPLC system (Shimadzu Nexera). Adequate separation of all compounds was achieved in nine minutes. (Figure 2 chromatogram shows the separation of all peaks). For this study, Shimadzu evaluated 1984 different instrument settings, and 6 different column and gradient combinations, to achieve excellent peak shape and resolution between peaks, as well as to maximize the signal-to-noise ratio of PFOA, PFHxS, PFNA, and PFOS. Mass spectrometry was performed on a Shimadzu LCMS- 8060NX with heated electrospray ionization operated in negative mode. Specific compound MRM transitions and associated internal standards are listed in Table 2. Chromatography was adjusted to provide sufficient separation of PFOA from potential cholic acid interferences, and to provide baseline resolution of branched and linear isomers (Figure 3). Perfluorotridecanoic acid PFTrDA 72629-94-8 0.1 Perfluorotetradecanoic acid PFTeDA 376-06-7 0.1 Perfluorobutanesulfonic acid PFBS 375-73-5 0.1 Perfluoropentansulfonic acid PFPeS 2706-91-4 0.1 Perfluorohexanesulfonic acid PFHxS 355-46-4 0.1 Perfluoroheptanesulfonic acid PFHpS 375-92-8 0.1 Perfluorooctanesulfonic acid PFOS 1763-23-1 0.1 Perfluorononanesulfonic acid PFNS 68259-12-1 0.1 Perfluorodecanesulfonic acid PFDS 335-77-3 0.1 Perfluoroundecanesulfonic acid PFUnDS 749786-16-1 0.1 Perfluorododecanesulfonic acid PFDoS 79780-39-5 0.1 Perfluorotridecanesulfonic acid PFTrDS 791563-89-8 0.1 Perfluorooctanesulfonamide PFOSA 754-91-6 0.1 9-Chlorohexadecafluoro-3-oxanonane-1-sulfonic acid 9Cl-PF3ONS 756426-58-1 0.1 11-Chloroeicosafluoro-3-oxaundecane-1-sulfonic acid 11Cl-PF3OUdS 763051-92-9 0.1 Hexafluoropropylene oxide dimer acid HFPO-DA 13252-13-6 0.1 4,8-Dioxa-3H-perfluorononanoic acid DONA 919005-14-4 0.1 1H,1H, 2H, 2H-Perfluorohexane sulfonic acid 4:2 FTS 757124-72-4 0.1 1H,1H, 2H, 2H-Perfluorooctane sulfonic acid 6:2 FTS 27619-97-2 0.1 1H,1H, 2H, 2H-Perfluorodecane sulfonic acid 8:2 FTS 39108-34-4 0.1 1H,1H, 2H, 2H-Perfluorododecane sulfonic acid 10:2 FTS 120226-60-0 0.1
Analyte Quantitation Ion Qualifier Ion Internal Standard PFBA 212.9 > 168.8 13 C 4 -PFBA PFPeA 262.9 > 218.8 13 C 5 -PFPeA PFHxA 313 > 268.8 313 > 118.9 13 C 5 -PFHxA PFHpA 362.9 > 318.9 362.9 > 168.8 13 C 4 -PFHpA PFOA 412.9 > 368.9 412.9 > 168.8 13 C 8 -PFOA PFNA 462.9 > 418.5 462.9 > 218.6 13 C 9 -PFNA PFDA 512.9 > 468.95 512.9 > 268.55 13 C 6 -PFDA PFUnA 562.9 > 518.95 562.9 > 269 13 C 7 -PFUnA PFDoA 612.9 > 268.6 612.9 > 168.6 13 C 7 -PFUnA PFTrDA 663 > 618.6 663 > 168.5 13 C 7 -PFUnA PFTeDA 713 > 669 713 > 168.5 13 C 7 -PFUnA PFBS 298.8 > 79.8 298.8 > 98.9 13 C 3 -PFBS PFPeS 348.8 > 79.8 348.8 > 98.9 13 C 4 -PFHpA PFHxS 398.8 > 79.8 398.8 > 98.9 13 C 3 -PFHxS PFHpS 448.8 > 79.8 448.8 > 99 13 C 6 -PFDA PFOS 498.8 > 79.8 498.8 > 98.9 13 C 8 -PFOS PFNS 548.9 > 79.8 548.8 > 99 13 C 7 -PFUnA PFDS 598.8 > 99 598.8 > 79.8 13 C 7 -PFUnA PFUnDS 649 > 99 649 > 80 13 C 8 -PFOS PFDoS 699 > 80 699 > 99 13 C 2 -PFDoA PFTrDS 749 > 80 749 > 279.6 13 C 7 -PFUnA PFOSA 498 > 78 498 > 477.95 13 C 8 -PFOSA 9Cl-PF3ONS 530.8 > 350.8 532.9 > 352.95 13 C 8 -PFOS 11Cl-PF3OUdS 631 > 451 632.9 > 452.95 13 C 7 -PFUnA HFPO-DA 284.8 > 168.8 284.8 > 118.9 13 C 3 -HFPO-DA DONA 376.9 > 250.8 376.9 > 84.9 13 C 4 -PFHpA 4:2 FTS 326.9 > 80.9 326.9 > 306.8 13 C 2 -4:2 FTS 6:2 FTS 426.9 > 80.8 426.9 > 406.9 13 C 2 -6:2FTS 8:2 FTS 526.9 > 506.45 526.9 > 80.55 13 C 3 -PFHxS 10:2 FTS 626.9 > 606.7 626.9 > 80.9 13 C 7 -PFUnA Figure 2: Chromatogram of 1.0 ng/g PFAS in a fish matrix with separation of all peaks in nine minutes Figure 3: Separation of PFOS from Cholic Acids and baseline resolution between PFOS branched and linear isomers Table 2: MRM Transitions and Internal Standard Associations 2.0 2.5 3.0 3.5 4.0 4.5 5.0 5.5 6.0 6.5 7.0 7.5 8.0 8.5 min 0 100000 200000 300000 400000 500000 600000 700000 800000 900000 1000000 1100000 1200000 1300000
■ Quantitative Analysis Calibration standards were processed the same as samples. A linear model not forced through zero isotopic dilution calibration in matrix-matched standards provided the best fit and best recoveries of analytes. Residuals of each point in the curve were ±25% of the expected value. Calibration curves for PFOA, PFHxS, PFNA, and PFOS are shown in Figures 3 – 6 respectively. Branched and linear isomers of PFHXS and PFOS were integrated together. Exact labeled analogs were used as isotope dilution standards, except in a few cases where we found that 13 C 2 suffered matrix interferences. In these cases, another non- interfering isotope was chosen. The analytes tested and the associated calibration isotopes are shown in Table 3. Figure 3: PFOA Calibration Curve Figure 4: PFHxS Calibration Curve Figure 5: PFNA Calibration Curve Figure 6: PFOS Calibration Curve Blank matrixes and three different concentrations ranging from the SMPR required LOQ to 50 times the estimated LOQ were analyzed in triplicate. Recovery and repeatability for each analyte at each concentration are given in Table 3. The LOQ for each analyte was estimated by spiking at concentrations at, or below, the required LOQs listed in SMPR-2023_003. The spiked samples were analyzed in triplicate and the mean and repeatability standard deviation were calculated. Then, the standard deviation was divided by the mean and multiplied by 100% to calculate the repeatability percent relative standard deviation (RSD). The LOQs for all matrices and compounds were determined using an Excel worksheet that compared each of the requirements of the SMPR including retention time, recovery, repeatability, S/N > 3 for the qualifier ion and an ion ratio of ±30%. PFBA, PFPeA, and PFOSA LOQ were set at the minimum concentration, meeting recovery and repeatability requirements and a S/N > 10. The lowest concentration to meet all the requirements of the SMPR was set as the LOQ. Figure 7 shows examples of the LOQ spike for PFHxS, PFNA, PFOA, and PFOS and their corresponding internal standards. Conc. (ng/g) %Accuracy 0.05 100 0.1 98.6 0.5 105 1 97.0 5 98.8 Conc. (ng/g) %Accuracy 0.05 94.6 0.1 110 0.5 105 1 97.7 5 92.9 Conc. (ng/g) %Accuracy 0.05 102 0.1 95.6 0.5 106 1 95.9 5 100 Conc. (ng/g) %Accuracy 0.05 99.9 0.1 99.2 0.5 107 1 96.2 5 97.7
Table 3: Recovery and repeatability for each analyte at each spike concentration Analyte Spike conc. (ppb) Average conc. (ppb) Standard Deviation %RSD Average Recovery% PFBA Blank 0.004 0.1 0.103 2.51 2.43 103.4 1 0.970 0.21 0.21 97.0 5 4.830 0.79 0.82 96.6 PFPeA Blank ND 0.1 0.096 2.11 2.20 96.2 1 0.976 0.46 0.47 97.6 5 4.920 0.61 0.62 98.4 PFHxA Blank 0.002 0.1 0.099 1.64 1.66 98.8 1 0.957 0.55 0.58 95.7 5 4.883 1.72 1.77 97.6 PFHpA Blank ND 0.1 0.100 0.44 0.44 100.1 1 0.980 1.72 1.76 98.0 5 4.998 1.42 1.42 100.0 PFOA Blank ND 0.1 0.100 1.50 1.51 99.6 1 0.960 1.43 1.49 96.0 5 4.852 0.81 0.84 97.0 PFNA Blank ND 0.1 0.101 2.74 2.72 100.9 1 0.984 1.11 1.12 98.4 5 5.039 2.25 2.24 100.8 PFDA Blank 0.000 0.1 0.098 3.36 3.42 98.2 1 0.936 1.39 1.48 93.6 5 4.800 0.57 0.59 96.0 PFUnA Blank 0.002 0.1 0.100 1.65 1.65 99.8 1 0.932 1.44 1.55 93.2 5 4.896 1.63 1.66 97.9 PFDoA Blank ND 0.1 0.093 10.48 11.29 92.9 1 0.939 2.17 2.31 93.9 5 4.833 1.42 1.47 96.6 PFTrDA Blank 0.004 0.1 0.107 8.56 8.05 106.4 1 0.996 3.93 3.95 99.6 5 4.889 4.60 4.71 97.8 PFTeDA Blank 0.009 0.1 0.114 8.25 7.23 114.1 1 1.029 12.47 12.12 102.9 5 4.860 5.60 5.76 97.2 PFBS Blank ND 0.1 0.100 2.77 2.78 99.7 1 0.945 1.76 1.86 94.5 5 4.836 1.26 1.30 96.7 PFPeS Blank ND 0.1 0.092 9.10 9.85 92.4 1 0.990 2.20 2.22 99.0 5 4.857 1.96 2.01 97.2 PFHxS Blank 0.002 0.1 0.105 6.51 6.21 104.9 1 0.968 2.02 2.09 96.8 5 4.838 3.11 3.21 96.7 PFHpS Blank 0.001 0.1 0.099 3.85 3.88 99.1 1 0.925 2.29 2.47 92.5 5 4.613 0.35 0.38 92.2
PFOS Blank 0.003 0.1 0.102 1.89 1.85 102.2 1 0.968 0.60 0.62 96.8 5 4.834 0.90 0.93 96.7 PFNS Blank ND 0.1 0.101 6.29 6.26 100.4 1 0.929 2.48 2.68 92.9 5 4.840 1.61 1.66 96.8 PFDS Blank ND 0.1 0.097 4.10 4.24 96.7 1 0.982 2.63 2.68 98.2 5 4.877 3.65 3.74 97.6 PFUnDS Blank 0.000 0.1 0.106 13.25 12.58 105.3 1 1.054 7.89 7.49 105.4 5 4.759 6.56 6.90 95.2 PFDoS Blank 0.005 0.1 0.114 16.66 14.59 114.1 1 1.084 20.30 18.72 108.4 5 4.694 2.70 2.88 93.9 PFTrDS Blank 0.008 0.1 0.119 14.78 12.41 119.1 1 1.113 24.60 22.11 111.3 5 4.911 7.02 7.15 98.2 PFOSA Blank 0.000 0.1 0.096 2.16 2.25 96.0 1 0.945 0.97 1.03 94.5 5 4.898 0.67 0.68 98.0 9Cl-PF3ONS Blank ND 0.1 0.097 1.14 1.17 97.2 1 0.965 0.44 0.45 96.5 5 4.901 1.82 1.86 98.0 11Cl-PF3OUdS Blank ND 0.1 0.097 6.63 6.82 97.3 1 0.943 3.67 3.89 94.3 5 4.796 3.06 3.20 95.9 HFPO-DA Blank 0.004 0.1 0.102 6.03 5.91 101.9 1 0.964 1.36 1.41 96.4 5 4.778 0.21 0.22 95.5 DONA Blank 0.001 0.1 0.100 1.42 1.43 99.6 1 0.956 0.90 0.94 95.6 5 4.974 1.10 1.11 99.5 4:2 FTS Blank 0.000 0.1 0.099 3.73 3.77 98.9 1 0.955 1.62 1.69 95.5 5 4.718 1.86 1.97 94.4 6:2 FTS Blank 0.001 0.1 0.098 2.91 2.98 97.7 1 0.944 1.03 1.09 94.4 5 4.664 1.00 1.07 93.3 8:2 FTS Blank 0.010 0.1 0.107 4.31 4.02 107.2 1 1.096 9.63 8.79 109.6 5 5.041 19.12 18.97 100.8 10:2 FTS Blank 0.008 0.1 0.110 5.70 5.21 109.4 1 1.029 7.31 7.10 102.9 5 4.028 8.96 11.12 80.6 ND = average results less than zero
Figure 7: LOQ Peaks with Internal Standards ■ Conclusion The Shimadzu LCMS-8060NX Triple Quadrupole Mass Spectrometer coupled with a Shimadzu Nexera Liquid Chromatograph was used in a single laboratory study to measure 30 PFAS compounds in a fish matrix and compared to criteria set by AOAC SMPR 2023.003. Chromatography conditions and the mass spectrometer were optimized to achieve excellent separation of all analytes, baseline resolution between linear and branched isomers, and a two- minute separation between PFOS and potentially interfering cholic acids. Precision and recovery and the experimentally determined LOQ are well within the requirements of the SMPR. ■ Reference 1) AOAC SMPR 2023.003 4.5 5.0 0 25000 50000 19:412.90>168.80(-) 19:412.90>368.90(-) 4.5 5.0 0 1000000 20:421.00>375.90(-) 5.0 5.5 0 2500 5000 7500 21:398.80>98.90(-) 21:398.80>79.80(-) 5.0 5.5 0 100000 200000 22:402.00>79.80(-) 5.0 5.5 0 25000 50000 23:462.90>218.60(-) 23:462.90>418.50(-) 5.0 5.5 0 500000 24:472.00>426.90(-) PFOA PFHxS PFNA PFOS 6.0 6.5 0 5000 10000 31:498.80>98.90(-) 31:498.80>79.80(-) 6.0 6.5 0 250000 30:507.00>79.80(-)
Copyright © 2024 Shimadzu Corporation and/or its affiliates. All rights reserved. SHIMADZU Scientific Instruments Inc. www.ssi.shimadzu.com Shimadzu Corporation www.shimadzu.com/an/ Founded in 1875, Shimadzu Corporation, a leader in the development of advanced technologies, has a distinguished history of innovation built on the foundation of contributing to society through science and technology. Established in 1975, Shimadzu Scientific Instruments (SSI), the American subsidiary of Shimadzu Corporation, provides a comprehensive range of analytical solutions to laboratories throughout North, Central, and parts of South America. SSI maintains a network of ten regional offices strategically located across the United States, with experienced technical specialists, service and sales engineers situated throughout the country, as well as applications laboratories on both coasts. For information about Shimadzu Scientific Instruments and to contact your local office, please visit our Web site at www.ssi.shimadzu.com LCMS-8050RX LCMS-8060RX LCMS-2020 Q-TOF LCMS-9030/9050 For Research Use Only. Not for use in diagnostic procedures. This publication may contain references to products that are not available in your country. Please contact us to check the availability of these products in your country. The content of this publication shall not be reproduced, altered or sold for any commercial purpose without the written approval of Shimadzu. See http://www.shimadzu.com/about/trademarks/index.html for details. Third party trademarks and trade names may be used in this publication to refer to either the entities or their products/services, whether or not they are used with trademark symbol “TM” or “”. Shimadzu disclaims any proprietary interest in trademarks and trade names other than its own. The information contained herein is provided to you "as is" without warranty of any kind including without limitation warranties as to its accuracy or completeness. Shimadzu does not assume any responsibility or liability for any damage, whether direct or indirect, relating to the use of this publication. This publication is based upon the information available to Shimadzu on or before the date of publication, and subject to change without notice. LCMS-8045RX LCMS-2050 02-SSI-LCMS-167 First Edition: September 2024
Please fill out the survey Related Products Some products may be updated to newer models. LCMS-TQ RX Series Triple Quadrupole LC- MS/MS Triple Quadrupole LC-MS/MS Related Solutions Food and Beverages Price Inquiry Product Inquiry Technical Service / Support Inquiry Other Inquiry Powered by TCPDF (www.tcpdf.org)