Application
News
www.shimadzu.com/an/
Shimadzu Corporation
Copyright © 2026 Shimadzu Corporation and/or its affiliates. All rights reserved.
For Research Use Only. Not for use in diagnostic procedures.
This publication may contain references to products that are not available in your country. Please contact us to check the availability of these
products in your country.
The content of this publication shall not be reproduced, altered or sold for any commercial purpose without the written approval of Shimadzu.
See http://www.shimadzu.com/about/trademarks/index.html for details.
Third party trademarks and trade names may be used in this publication to refer to either the entities or their products/services, whether or not
they are used with trademark symbol “TM” or “”.
Shimadzu disclaims any proprietary interest in trademarks and trade names other than its own.
The information contained herein is provided to you "as is" without warranty of any kind including without limitation warranties as to its
accuracy or completeness. Shimadzu does not assume any responsibility or liability for any damage, whether direct or indirect, relating to the
use of this publication. This publication is based upon the information available to Shimadzu on or before the date of publication, and subject
to change without notice.
Kratos Analytical Ltd.
www.kratos.com
First Edition: May 2026 12-MO-553-EN
FlexiFocus is a trademark of Shimadzu Corporation or its affiliated companies in Japan and/or other countries.
We have shown that it is feasible to implement a fast and
simple screening protocol using MALDI-TOF mass spectrometry
following the solid phase extraction step for the analysis of
PFAS samples, to prevent contamination of instrument
hardware in LC-MS/MS workflows.
The use of FlexiFocus slides permits deposition of analytes and
matrix primarily composed of methanol. This improves the
precision of concentration estimates in comparison with
aqueous preparations, although the lower limit for the
concentration estimate remains unchanged due to the loss of
linearity in the calibration curve.
For water samples, limited solubility introduces uncertainty in
concentrations determined by MALDI-TOF mass spectrometry .
Therefore, samples should be screened after solid-phase
extraction when the samples are reconstituted primarily in
methanol.
Whilst limits of detection are lower with the use of FlexiFocus
slides, the loss of linearity in the lower concentration range
prevents the implementation of screening protocols below
approximately 500 ppt for PFOS and 5 ppb for PFOA.
<References>
1. Detection of PFAS in Aqueous Samples by Matrix Assisted
Laser Desorption IonisationTime-of-Flight (MALDI-TOF) Mass
Spectrometry Application News No. 12-MO-543-EN
2. Tighe, M et al, Rapid PFAS removal from water with
floating polymer assisted by air bubbles, Chemosphere 377,
2025.
Conclusions
The calibration curve for PFOA was then recalculated across the
range of 5–100 ppb (Fig. 10). Representative spectra are shown
in Fig. 11. The R
2
value of 0.99 across this range again allows for
a good level of confidence in the correct identification of those
samples which could result in instrument contamination.
Fig. 10 Calibration Curve for PFOA (5–100 ppb) in 0.5 mg/mL TMGN
Fig. 11 Representative spectra for PFOA (5–100 ppb) in 0.5 mg/mL TMGN
Analysis of PFOA
PFOA was selected for analysis to provide further evidence for
the suitability of this method. PFOA samples were prepared
across a 1–100 ppb range in 100% methanol. The samples were
premixed with a 0.5 mg/mL TMGN solution (100% methanol)
and 1 µL was spotted onto a 96 well FlexiFocus slide. When
calculating the calibration curve, it was apparent that, at
concentrations below 5 ppb, linearity was lost (Fig. 9).
Fig. 9 Calibration Curve for PFOA (1–10 ppb) in 0.5 mg/mL TMGN
R
2
= 0.9927
R
2
= 0.6608