User Benefits
Application
News
LabSolutions Detect automates impurity peak detection and identification, improving the efficiency of impurity management.
By coupling a single quadrupole mass spectrometer (LCMS-2050), mass information can be acquired as qualitative information
for impurity peaks.
Streamlining Impurity Management Using
LabSolutions Detect
Shinichi Fujisaki
Anomaly Peak Detection Support Software
In the pharmaceutical development process, the qualitative and
quantitative analysis and appropriate management of impurities
are essential to ensure product safety. For example, during active
pharmaceutical ingredient (API) synthesis, various impurities are
generated throughout multi-step synthetic routes from starting
materials to the final API. Accordingly, the ICH guidelines¹⁾ define
threshold levels at which impurity identification is required, and
such impurities must be controlled within acceptable limits. In
addition, stability testing of pharmaceuticals is conducted to
evaluate time-dependent changes under various exposure
conditions in order to ensure product safety and determine shelf
life. Degradation products generated under these exposure
conditions must also be appropriately managed to monitor such
changes over time. However, when multiple impurities and
degradation products are managed manually, tasks such as peak
detection and peak identification carry the risk of missed
detections and human error. LabSolutions Detect, an anomaly
peak detection support software, automates peak detection and
peak identification for arbitrary peaks, including impurities and
degradation products. In this article, case examples are presented
in which peak detection and peak tracking were automated for
multiple model samples assuming different impurity profiles,
thereby streamlining impurity management. Furthermore, the
combined use of a single quadrupole mass spectrometer (LCMS-
2050) demonstrates that mass information can be acquired as
qualitative information for each peak.
Introduction
Fig. 2 shows the results of automatic impurity detection and
identification performed by LabSolutions Detect in accordance
with the criteria set as detection thresholds (Fig. 1). The blue-
framed area in Fig. 2 shows the results for the control, in which
three known compounds (Known 1–3) were detected. The red-
framed area in Fig. 2 shows the results for Samples 1–5, where the
area percentages of the detected impurities are automatically
entered as red values. In addition, because peak identification is
also automated, identical impurities are displayed in the same
column. For example, in Sample 2, two impurities (New 2 and
New 3) were detected, and New 2 was also detected in Samples 3
and 4. Furthermore, for the three known compounds (Known 1–
3), the area percentages were automatically entered for Samples
1–5, indicating that these compounds were detected in all
samples. Fig. 3 shows the individual chromatograms for Samples
1–5, in which automatically detected impurity peaks are
highlighted in red, facilitating the identification of unknown
impurities. In addition, a comparison of the chromatograms for
Samples 1–5 is shown in Fig. 4. By combining these results with
those shown in Fig. 2, the number of impurities detected across
multiple chromatograms and their elution times can be visually
confirmed. Moreover, mass information obtained using the
LCMS-2050 can be utilized as qualitative information for each
impurity.
Streamlining Impurity Management by
LabSolutions Detect
Ultra High Performance Liquid Chromatograph
As target compounds, a control sample free of impurities and five
model samples assuming different impurity profiles were used;
the model samples were artificially prepared as mixtures of
different low-molecular-weight pharmaceuticals. The analytical
conditions are summarized in Table 1. LabSolutions Detect
automatically performs peak detection according to predefined
criteria. In this study, the upper limit of the area percentage for
impurity peaks was set to 0.1%, and peak detection and peak
tracking were automatically performed for Samples 1–5 (Table 1).
Target Compounds
Table 1 Analytical Conditions
Fig. 1 Criteria for Unknown Impurities
Control : Benzoic acid, Naproxen, Probenecid
Sample 1 : Control with added quinidine
Sample 2 : Control with added antipyrine, hydrocortisone
Sample 3 : Control with added antipyrine, furosemide
Sample 4 : Control with added antipyrine, indometacin
Sample 5 : Control with added indometacin
Mobile phase
Pump A : 0.1% formic acid in water
Pump B : Acetonitrile
Column : Shim-pack Scepter C18-120
*1
(100 mm × 3.0 mm I.D., 1.9 µm)
Injection Vol. : 0.5 µL
*1 P/N : 227-31013-03
Criteria of peak detection : Upper limit of impurity (area %) : 0.1
LC Conditions
B Conc. : 5% (0 min)→90% (5 min)→5% (5-10 min)
Column Temp. : 40 °C
Flow rate : 0.7 mL/min
Detection : 254 nm (SPD-M40, STD cell)
System : LCMS-2050
Ionization : ESI/APCI (DUIS), positive and negative mode
Mode : SCAN (m/z 80-1000)
Nebulizing gas flow : 2.0 L/min
Drying gas flow : 5.0 L/min
Heating gas flow : 7.0 L/min
DL Temp. : 200 °C
Desolvation Temp. : 450 °C
Interface voltage : +3.0 kV / -2.0 kV
MS Conditions
System : Nexera X3