User Benefits Application News Using narrow-bore columns enables a substantial reduction in solvent consumption. Nexera X4 features a low-dispersion design with minimized internal volume, enabling sharp peak shapes even when using narrow-bore columns. Using LabSolutions MD enables a more efficient workflow for developing and optimizing separation conditions. Achieving Solvent Reduction and Lower Running Costs with Nexera X4 Shinichi Fujisaki In recent years, there has been growing interest in sustainability within analytical laboratories. In liquid chromatography, the large volumes of organic solvents used as mobile phases present challenges from both environmental and cost perspectives. One approach to reducing solvent consumption is to downscale analyses by using small-diameter columns. Such solvent-saving strategies not only contribute substantially to lowering running costs, but also reduce the amount of waste solvent, enabling more environmentally conscious laboratory operations. However, analytical downscaling requires effective suppression of band broadening within flow channels and highly accurate low-volume injections. Nexera X4 (Fig. 1) delivers exceptional instrument performance that meets these demands and provides stable analytical results even when small-diameter columns are used. This article presents a case study in which analysis was downscaled and solvent consumption was significantly reduced by combining Nexera X4 with a 1-mm-ID column (Shim-pack NovaCore C18-HB). In addition, the optimization of separation conditions was streamlined by using LabSolutions MD, a dedicated software for supporting method development. Introduction Fig. 2 shows the chromatogram obtained under the analytical conditions listed in Table 1 (gradient time: 2 min, column oven temperature: 40 ˚C). Compounds (C) and (D) co-eluted (red circle in Fig. 2), indicating that optimization of the separation is required. LabSolutions MD enables anyone to perform rapid and error-free exploration of separation conditions by automatically generating analysis schedules under comprehensive combinations of parameters (steps (1)–(4) in Fig. 3). For example, when varying the gradient time (red box in Fig. 3), an entered center value of 3 min is adjusted by one step above and below the center at 1-min increments, resulting in automatically generated schedules of 2, 3, and 4 min. In this study, nine analytical schedules (Fig. 4) were automatically generated by varying the gradient time (2, 3, 4 min; three patterns) and column oven temperature (30, 40, 50 ˚C; three patterns), enabling comprehensive evaluation of the separation. Enhancing the Efficiency of Separation Optimization with LabSolutions MD The analytical conditions and target compounds are summarized in Table 1. In this study, a mixed standard containing eight small- molecule pharmaceuticals was used as a model sample. Separation conditions for this sample were optimized using a 1- mm-ID column. Analytical Conditions and Target Compounds Software for Efficient Method Development High Performance Liquid Chromatograph 1 2 min 0 0.4 AU Gradient time 2 min/Oven Temp. 40 °C (A) (B) (C) (D) (E) (F) (G) (H) Fig. 1 Nexera X4 Fig. 2 Chromatogram Obtained Using the 1-mm-ID Column (Before separation optimization) (Peak labels correspond to those in Table 1) Table 1 Analytical Conditions and Target Compounds *1:P/N 227-32901-01 *2:X (Gradient time)= 2, 3, 4 Analytical conditions B Conc. :5%(0 min)→70%(X *2 min)→5%(X~X+5 min) Column Temp. :30, 40, 50 ˚C Flow rate :0.1 mL/min Mixer :Micro mixer Sample loop Vol. :15 µL Injection Vol. :0.3 µL Detection :254 nm (SPD-M40 X4, STD cell) Mobile phase Pump A : 0.1% formic acid in water Pump B : Acetonitrile Sample : (A) Antipyrine, (B) Benzoic acid, (C) Salicylic acid, (D) Hydrocortisone, (E) Furosemide, (F) Naproxen, (G) Probenecid, (H) Diclofenac System : Nexera X4 Column : Shim-pack NovaCore C18-HB *1 (50 mm × 1.0 mm I.D., 1.7 µm)
Application News 0% 50% 100% 1mm 2.1mm 3mm Solvent Consumption (2) Set center value (4) Set step number (3) Set step size (1) Select the parameters (1) (2) (3) (6) (5) (4) (9) (8) (7) (9) Gradient time 4 min/Oven Temp. 50 ℃ (8) Gradient time 3 min/Oven Temp. 50 ℃ (7) Gradient time 2 min/Oven Temp. 50 ℃ (6) Gradient time 4 min/Oven Temp. 40 ℃ (5) Gradient time 3 min/Oven Temp. 40 ℃ (4) Gradient time 2 min/Oven Temp. 40 ℃ (3) Gradient time 4 min/Oven Temp. 30 ℃ (2) Gradient time 3 min/Oven Temp. 30 ℃ (1) Gradient time 2 min/Oven Temp. 30 ℃ 1 2 min 3 0 0.5 AU 1 2 min 3 0 0.4 AU Gradient time 3 min/Oven Temp. 50 °C (A) (B) (C) (D) (E) (F) (H) (G) Fig. 7 Comparison of Column Inner Diameter and Solvent Consumption (Linear velocity kept constant across column IDs) (Using the 3 mm column’s solvent consumption as 100% for comparison) 11% 49% 100% Fig. 5 Chromatograms Obtained by Varying the Gradient Time and Column Oven Temperature Fig. 6 Chromatogram Under the Optimal Separation Conditions: (8) in Fig. 5 (Peak labels correspond to those in Table 1) Fig. 7 compares solvent consumption for column inner diameters of 1.0, 2.1, and 3.0 mm. Switching from a 3-mm column to a 1-mm column reduces solvent consumption by approximately 9-fold, while switching from a 2.1-mm column to a 1-mm column achieves a reduction of about 4.5-fold. These reductions not only help lower running costs but also contribute to reducing environmental impact by decreasing the volume of waste solvent. Reduction of Solvent Consumption Using a 1-mm Column Fig. 5 shows the chromatograms obtained by comprehensively varying the gradient time and column oven temperature. A longer gradient time tended to improve the separation between peaks, while higher column oven temperatures resulted in earlier elution of the compounds and a shorter overall analysis time. Considering both the resolution between compounds and the need to minimize analysis time, the condition of a 3-min gradient and a column oven temperature of 50 ˚C ((8) in Fig. 5) was selected as the optimal separation condition. Fig. 6 shows the chromatogram obtained under the optimal conditions. Fig. 3 Analysis Schedule Creation Screen Fig. 4 Automatic Generation of Analysis Schedule
Application News www.shimadzu.com/an/ Shimadzu Corporation © Shimadzu Corporation, 2026 For Research Use Only. Not for use in diagnostic procedures. This publication may contain references to products that are not available in your country. Please contact us to check the availability of these products in your country. The content of this publication shall not be reproduced, altered or sold for any commercial purpose without the written approval of Shimadzu. See http://www.shimadzu.com/about/trademarks/index.html for details. Third party trademarks and trade names may be used in this publication to refer to either the entities or their products/services, whether or not they are used with trademark symbol “TM” or “”. Shimadzu disclaims any proprietary interest in trademarks and trade names other than its own. The information contained herein is provided to you "as is" without warranty of any kind including without limitation warranties as to its accuracy or completeness. Shimadzu does not assume any responsibility or liability for any damage, whether direct or indirect, relating to the use of this publication. This publication is based upon the information available to Shimadzu on or before the date of publication, and subject to change without notice. First Edition: Jun. 2026 01-01174-EN Nexera, LabSolutions and Shim-pack are trademarks of Shimadzu Corporation or its affiliated companies in Japan and/or other countries. This article introduced an approach for reducing solvent consumption by using small-diameter columns. Reducing solvent usage not only lowers running costs but also contributes to reducing environmental impact by decreasing the amount of waste solvent. Although suppressing dispersion within the flow path is essential when scaling down an analysis, Nexera X4 maximizes the performance of small-diameter columns through its low-dispersion design. Furthermore, when combined with the method-development support software LabSolutions MD, it also streamlines the exploration of optimal separation conditions. Conclusion 1 3 min 0 0.2 AU (2) Typical UHPLC System_Gradient time 3 min/Oven Temp. 50 ℃ (1) Nexera X4_Gradient time 3 min/Oven Temp. 50 ℃ (1) (2) 0.4 5 (A) (B) (D) (E) (F) (H) (G) (C) (A) (B) (D) (E) (F) (H) (G) (C) Fig. 8 Comparison of Chromatograms Under the Optimal Conditions (1) Nexera X4, (2) Typical UHPLC System (Peak labels correspond to those in Table 1) When using 1-mm-ID columns, minimizing extra-column dispersion caused by system flow channel is essential for maximizing the inherent performance of the column. Fig. 8 compares chromatograms obtained with Nexera X4 and a typical UHPLC system using a 1-mm-ID column. While Nexera X4 provides sharp peak shapes, the typical UHPLC system shows broader peaks due to the influence of extra-column dispersion. With its low-dispersion design and minimized internal volume, Nexera X4 enables full performance of small-diameter columns. Maximizing the Performance of 1-mm-ID Columns with Nexera X4
Please fill out the survey Related Products Some products may be updated to newer models. Nexera X4 Ultra High-Performance Liquid Chrom… Ultra High-Performance Liquid Chromatograph Related Solutions Pharmaceutical and Biopharmaceutical Small Molecule Pharmaceutical Price Inquiry Product Inquiry Technical Service / Support Inquiry Other Inquiry Powered by TCPDF (www.tcpdf.org)
Please fill out the survey Related Products Some products may be updated to newer models. Nexera X4 Ultra High-Performance Liquid Chrom… Ultra High-Performance Liquid Chromatograph Method Development System Automatic Optimization of Gradient Conditions with… Shim-pack NovaCore Organic Silica Hybrid Core-Shell Column Related Solutions Pharmaceutical and Biopharmaceutical Small Molecule Pharmaceutical Price Inquiry Product Inquiry Technical Service / Support Inquiry Other Inquiry Powered by TCPDF (www.tcpdf.org)