User Benefits
Application
News
Using narrow-bore columns enables a substantial reduction in solvent consumption.
Nexera X4 features a low-dispersion design with minimized internal volume, enabling sharp peak shapes even when using
narrow-bore columns.
Using LabSolutions MD enables a more efficient workflow for developing and optimizing separation conditions.
Achieving Solvent Reduction and Lower Running
Costs with Nexera X4
Shinichi Fujisaki
In recent years, there has been growing interest in sustainability
within analytical laboratories. In liquid chromatography, the large
volumes of organic solvents used as mobile phases present
challenges from both environmental and cost perspectives. One
approach to reducing solvent consumption is to downscale
analyses by using small-diameter columns. Such solvent-saving
strategies not only contribute substantially to lowering running
costs, but also reduce the amount of waste solvent, enabling
more environmentally conscious laboratory operations. However,
analytical downscaling requires effective suppression of band
broadening within flow channels and highly accurate low-volume
injections. Nexera X4 (Fig. 1) delivers exceptional instrument
performance that meets these demands and provides stable
analytical results even when small-diameter columns are used.
This article presents a case study in which analysis was
downscaled and solvent consumption was significantly reduced
by combining Nexera X4 with a 1-mm-ID column (Shim-pack
NovaCore C18-HB). In addition, the optimization of separation
conditions was streamlined by using LabSolutions MD, a
dedicated software for supporting method development.
Introduction
Fig. 2 shows the chromatogram obtained under the analytical
conditions listed in Table 1 (gradient time: 2 min, column oven
temperature: 40 ˚C). Compounds (C) and (D) co-eluted (red circle
in Fig. 2), indicating that optimization of the separation is
required. LabSolutions MD enables anyone to perform rapid and
error-free exploration of separation conditions by automatically
generating analysis schedules under comprehensive
combinations of parameters (steps (1)–(4) in Fig. 3). For example,
when varying the gradient time (red box in Fig. 3), an entered
center value of 3 min is adjusted by one step above and below
the center at 1-min increments, resulting in automatically
generated schedules of 2, 3, and 4 min. In this study, nine
analytical schedules (Fig. 4) were automatically generated by
varying the gradient time (2, 3, 4 min; three patterns) and column
oven temperature (30, 40, 50 ˚C; three patterns), enabling
comprehensive evaluation of the separation.
Enhancing the Efficiency of Separation
Optimization with LabSolutions MD
The analytical conditions and target compounds are summarized
in Table 1. In this study, a mixed standard containing eight small-
molecule pharmaceuticals was used as a model sample.
Separation conditions for this sample were optimized using a 1-
mm-ID column.
Analytical Conditions and Target
Compounds
Software for Efficient Method Development
High Performance Liquid Chromatograph
1 2 min
0
0.4
AU
Gradient time 2 min/Oven Temp. 40 °C
(A)
(B)
(C)
(D)
(E)
(F)
(G)
(H)
Fig. 1 Nexera X4
Fig. 2 Chromatogram Obtained Using the 1-mm-ID Column
(Before separation optimization)
(Peak labels correspond to those in Table 1)
Table 1 Analytical Conditions and Target Compounds
*1:P/N 227-32901-01
*2:X (Gradient time)= 2, 3, 4
Analytical conditions
B Conc. :5%(0 min)→70%(X
*2
min)→5%(X~X+5 min)
Column Temp. :30, 40, 50 ˚C
Flow rate :0.1 mL/min
Mixer :Micro mixer
Sample loop Vol. :15 µL
Injection Vol. :0.3 µL
Detection :254 nm (SPD-M40 X4, STD cell)
Mobile phase
Pump A : 0.1% formic acid in water
Pump B : Acetonitrile
Sample : (A) Antipyrine, (B) Benzoic acid, (C) Salicylic acid,
(D) Hydrocortisone, (E) Furosemide, (F) Naproxen,
(G) Probenecid, (H) Diclofenac
System : Nexera X4
Column : Shim-pack NovaCore C18-HB
*1
(50 mm × 1.0 mm I.D., 1.7 µm)