TGA-50/50H And TGA-51/51H

Thermogravimetric Analyzer

  • Temperature range: Room temperature to 1000°C/Room temperature to 1500°C (H-type)
  • Measurement range: ±20, ±200 mg (50-type)
  • Measurement range: ±20, ±200, ±2000 mg (51-type)
  • Maximum sample weight: 1 g (including wind bag)/10 g (including wind bag, 51-type)
 
TGA-50
TGA-50H
TGA-51
TGA-51H
Temperature Range
Ambient to
1000°C
Ambient to
1500°C
Ambient to
1000°C
Ambient to
1500°C
Weight measuring range
20mg, 200mg
20mg, 200mg, 2g
Readability
0.001mg
Sample mass
1 g including tare
10g including tare
Programmable heating rate
0.1°C/hour to 99.9°C/min
0.1°C/hour to 50.0°C/min
0.1°C/hour to 99.9°C/min
0.1°C/hour to 50.0°C/min
Hold time
0 to 999min, 0 to 999hour
Temperature program format
99 steps maximum
Temperature program file
up to 100 files
Cooling method
Air-cooled
Signal output
Analog and digital
Atmosphere control
Built-in gas flowmeter (250ml/min maximum)
Dimensions
W173xD550xH500mm
W173xD600xH540mm
Power supply
AC100V, 120V, 220V, 240V, 500VA, 50/60Hz
AC100V, 120V, 220V, 240V, 1.2kVA, 50/60Hz
AC100V, 120V, 220V, 240V, 1kVA, 50/60Hz
AC100V, 120V, 220V, 240V, 1.5kVA, 50/60Hz

For Research Use Only. Not for use in diagnostic procedures.

This page may contain references to products that are not available in your country. Please contact us to check the availability of these products in your country.

Top of This Page