High-Sensitivity Analysis of Nonylphenol in River Water Using GC-MS/MS

Nonylphenol (NP) is used as a raw material for the production of surfactants, and as an antioxidant used to protect rubber and plastics, etc. However, in recent years, it has been specified as a substance that can cause endocrine disruption in the environment (environmental hormone). We investigated the use of a high m/z selectivity triple quadrupole gas chromatograph-mass spectrometer (GC-MS/MS) in the analysis of NP in river water, which typically contains many contaminants. Analysis was possible without adversely affecting identification accuracy, even when omitting the cleanup procedure that may reduce the recovery rate.

Content Type:
Application
Document Number:
LAAN-A-MS-E035
Product Type:
Mass Spectrometry, Gas Chromatograph-Mass Spectrometry
Keywords:
Waterworks law, Water quality, Water quality standard, Nonylphenol, Environmental analysis, Tap water, Environmental water, MS/MS, TQ, Environment, Drinking Water, Pool Water, GCMS-TQ8040
Language:
English
File Name:
jpo214100.pdf
File Size:
1,175kb

Free Download

For Research Use Only. Not for use in diagnostic procedures.

This page may contain references to products that are not available in your country. Please contact us to check the availability of these products in your country.