Determination of Microcystins in Drinking Water by Ultra High Performance Liquid Chromatography/Triple Quadrupole Mass Spectrometry
An ultra high performance liquid chromatography/electrospray ionization tandem mass spectrometry (UHPLC/ESI-MS/MS) method has been developed to quantify trace levels of ten MCs (MC-RR, Demethyl-RR, LR, LY, LW, LA, YR, WR, CF and NOD) in water. The calibration curves of ten MCs were established and the correlation coefficients were better than 0.999. The LODs of the 10 MCs ranged from 0.02 to 0.5 μg/L.
- Content Type:
- Paper
- Document Number:
- PO-CON1341E
- Product Type:
- Liquid Chromatography, Liquid Chromatograph-Mass Spectrometry, Mass Spectrometry
- Keywords:
- Drinking Water, Microcystins, Environment, Food and Beverages, Drinking Water, Pool Water, Food QAQC, LCMS-8040, Nexera UHPLC, LC-30AD