ICP-MS Features and Analytes
Basics of Inductively Coupled Plasma Mass Spectrometry
ICP-MS for Elemental Analysis
ICP-MS for elemental analysis is one of the most powerful and sensitive approaches available for multi-element and trace measurements. As a leading solution in ICP-MS elemental analysis, this technique enables accurate detection and quantification of elements at ppt to ppm levels across a wide range of sample types. Compared to other elemental analysis methods, ICP-MS offers exceptional sensitivity, wide dynamic range, and robust multi-element capability, making it ideal for trace elemental analysis, heavy metal analysis, and ultra-low level impurity detection. As one of the most advanced elemental analysis instruments, ICP-MS supports quantitative elemental analysis in environmental testing, pharmaceutical elemental analysis, semiconductor materials, metals, and high-purity chemicals. Its ability to manage matrix effects, minimize mass interferences, and deliver reliable detection limits makes ICP-MS a preferred choice for laboratories requiring high-sensitivity elemental analysis and comprehensive elemental characterization.
1. Analyte Elements
ICP-MS can analyze the approximately 70 elements highlighted in the periodic table below and can measure many of these elements at ppt levels.
Although sensitivity differs by element, ICP-MS tends to offer good sensitivity when measuring elements with the following three characteristics:
(1) A mass number with a high isotopic ratio
(2) An element with a high mass number (less interference)
(3) An element with low ionization energy
Because ICP-MS can also measure the Halogen elements (except fluorine), it can measure a broader range of elements than inductively coupled plasma atomic emission spectroscopy (ICP-AES) and atomic absorption (AA) spectroscopy.
| 1 | 2 | 3 | 4 | 5 | 6 | 7 | 8 | 9 | 10 | 11 | 12 | 13 | 14 | 15 | 16 | 17 | 18 |
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| H | He | ||||||||||||||||
| Li | Be | B | C | N | O | F | Ne | ||||||||||
| Na | Mg | AI | Si | P | S | CI | Ar | ||||||||||
| K | Ca | Sc | Ti | V | Cr | Mn | Fe | Co | Ni | Cu | Zn | Ga | Ge | As | Se | Br | Kr |
| Rb | Sr | Y | Zr | Nb | Mo | Tc | Ru | Rh | Pd | Ag | Cd | In | Sn | Sb | Te | I | Xe |
| Cs | Ba | La-Lu | Hf | Ta | W | Re | Os | Ir | Pt | Au | Hg | Tl | Pb | Bi | Po | At | Rn |
| Fr | Ra | Ac-Lr | |||||||||||||||
| La-Lu | La | Ce | Pr | Nd | Pm | Sm | Eu | Gd | Tb | Dy | Ho | Er | Tm | Yb | Lu | ||
| Ac-Lr | Ac | Th | Pa | U | Np | Pu | Am | Cm | Bk | Cf | Es | Fm | Md | No | Lr | ||
2. Comparing Elemental Analysis Methods
Table 1 below compares ICP-MS with other elemental analysis methods that measure liquid samples. Because AA measures a single element at a time, it is ideal for the low cost, routine analysis of a small number of elements. However, being a single-element method, AA cannot be used for qualitative analysis and analysis times increase in proportion with the number of elements that need to be measured.
ICP-AES(OES) and ICP-MS incur a greater initial cost compared to AA but are more efficient techniques as they can measure multi-element.
ICP-MS is generally more efficient than ICP-AES because it performs higher sensitivity measurements; however, because ions are introduced directly into the instrument, ICP-MS has lower matrix tolerance and more maintenance than ICP-AES. Accuracy can also be an issue when introducing a wide variety of ions into the instrument. Taking arsenic and selenium as examples, many fields of study require high-sensitivity analysis of these elements due to their toxicity. Both ICP-AES and AA require a pretreatment process based on hydride generation that is specific to each element to achieve these highly sensitive measurements, while ICP-MS can multi-element analysis both arsenic and selenium in addition to many other elements in a sample without complex pretreatment. However, like spectral interference in ICP-AES, mass interferences can occur in ICP-MS due to polyatomic clusters or mass (isobaric) overlap. Modern ICP-MS are specially designed to deal with these issues.
Table 1: Comparison of elemental analysis instrumentation
| AA flame | AA furnace | ICP-AES | ICP-MS | |
|---|---|---|---|---|
| Sensitivity | ppb to ppm | ppt to ppb | ppb | ppt |
| Concentration range | Narrow (around 2 digits) | Narrow (around 2 digits) | Wide (ppb to %) | Wide (ppt to ppm) |
| Matrix tolerance | Excellent | Average | Good | Average |
| Analysis speed (time per specimen) | 20 seconds (single-element analysis) |
5 minutes (single-element analysis) |
3 to 5 minutes (multi-element analysis) |
3 to 5 minutes (multi-element analysis) |
| Multi-element analysis | Not possible | Not possible | Possible (qualitative analysis possible) |
Possible (qualitative analysis possible) |
| Initial cost | Excellent | Good | Good | Average |
| Running costs, Gas used | Excellent Acetylene, air, (N2O) |
Good Ar |
Average Ar |
Average Ar, He, reaction gas (H2, NH3) |
| Ease of maintenance | Excellent | Average | Good | Average |
3. Samples Analyzable by ICP-MS
ICP-MS is typically used to analyze aqueous sample solutions, with the samples normally acidified to stabilize the analyte elements in solution with 1-2 % (v/v) nitric acid and Table 2 reviews some general guidelines. The addition of acid prevents element precipitation and absorption to the container. Hydrochloric acid is added for better solubilization and stabilization of elements like mercury and noble metals, though acids other than nitric acid cause spectral interference and their use should be limited when possible.
Solid samples require dissolution in solution to prepare them for ICP-MS analysis. Recently, microwave digestion systems have been used to prepare samples for ICP-MS analysis by performing acidic decomposition in sealed high-pressure vessels. This form of sample preparation is fast and effective and is becoming more common in analytical laboratories, especially for difficult to digest matrices. Hydrofluoric acid is used to decompose silicate samples and samples with strong oxide bonds. However, hydrofluoric acid is highly caustic and requires caution as it damages glass in the sample introduction system along with causing spectral interference. (If hydrofluoric acid must be used, a dedicated chemically resistant sample introduction system should be installed.)
Samples that contain sediment can clog the nebulizer and require filtering before introduction. Samples with a high salt concentration or large amounts of organic material can also clog the sample introduction system (due to salt deposition and soot generation in plasma), causing fluctuations in sensitivity. When the sample matrix and its effect on analysis is a concern, either the sample is diluted within the limits allowed by measurement sensitivity or the organic materials in the sample are decomposed.
ICP-MS can also analyze samples in organic solvents, but the sensitivity and accuracy of this analysis is lower than for aqueous solutions. Moreover, there is a risk of interference from carbon atoms. Analyzing samples in organic solvents also requires a sample introduction system designed for organic solvents with an additional Ar/O2 gas stream to reduce soot generation.
Table 2: General guidelines for good ICP-MS sample preparation
| Solvent | Water (or organic solvent) |
|---|---|
| Matrix |
|
| Acid |
|
| Sample volume required (minimum) |
|