Standard specifications
| Turbo molecular pump (Note 1) | TMP-V2804LM | |
|---|---|---|
| Inlet flange | VG250 | |
| Outlet flang | KF40 | |
| Cooling method | Water | |
| Ultimate pressure (after baking) | 10-7 Pa order | |
| Maximum allowable Argon gas flow rate (Note 2, Note 3) | 1500 mL/min | |
| Maximum allowable inlet pressure (N2 continuous exhaust) |
7 Pa | |
| Maximum allowable outlet pressure (N2 continuous exhaust) |
270 Pa | |
| Pumping speed (Note 4) | N2 | 2800 L/s |
| Ar | 2700 L/s | |
| Compression ratio | N2 | 1×109 or more |
| He | 1×105 | |
| H2 | 6×103 | |
| Rated speed | 27600 rpm | |
| Start-up time | 16 minutes or les | |
| Mounting position | In any desired direction | |
| Vibration level (by Shimadzu's method) | 0.01 μm or les(single amplitude) | |
| Recommended flow rate of purge gas (Note 2) | 30 mL/min | |
| Recommended pumping speed of backing pump in case of gas purge | 1500 L/min or more | |
| Environment temperature | 10 ~ 40 ℃ | |
| Admissible ambient magnetic field | Radial direction | 3 mT |
| Axial direction | 15 mT | |
| Water | Flow Rate | 3 ~ 4 L/min |
| Pressure | 0.2 ~ 0.4 MPa | |
| Temperature | 19 ~ 30 degrees C.(above dew point) | |
| Display | LED | POWER / STATUS / ROTATION / NET |
| Communication | Contact | REMOTE(D-sub 37 pin male, Screw lock size M2.6) |
| Input:START / STOP / RESET / LOW SPEED | ||
| Output:ROTATION / ACC. / BRAKE / NORMAL /REMOTE / ALARM / WARNING / CONNECTION | ||
| Serial | RS-232C / RS-485(D-sub 9 pin male, Screw lock size M2.6) | |
| Speed variation | Speed is variable between 25 % and 100 % of the rated speed. (set as 0.1 %) | |
| Input electric power | Voltage | Single phase AC 200 ~ 240 V(50 / 60 Hz) |
| Maximum power | 1.2 kVA | |
| Mass | 94 kg | |
- (Note 1)Do not use corrosive gas because there is possibility that the pump breaks down by the product material.
- (Note 2)mL/min: volume flow rate at 0 degrees C., 1 atm (Compatible with SCCM).
- (Note 3)The maximum allowable flow rate depends on the gas type, the APC valve temperature, the gas purge flow rate and outlet pressure.
- (Note 4)Without a protective net. Pumping speed for N2 is 2550 L/s with a protective net.
This page may contain references to products that are not available in your country. Please contact us to check the availability of these products in your country.