Standard specifications
| Turbo molecular pump (Note 1) | TMP-V3304LM (0) | TMP-V3304LM | |
|---|---|---|---|
| Inlet flange | VG300 | ISO320B / VG350 | |
| Outlet flange | KF40 | ||
| Cooling method | Water | ||
| Ultimate pressure (after baking) | 10-7 Pa order | ||
| Maximum allowable Argon gas flow rate (Note 2, Note 3) | 1500 mL/min | ||
| Maximum allowable inlet pressure (N2 continuous exhaust) |
7 Pa | ||
| Maximum allowable outlet pressure (N2 continuous exhaust) |
270 Pa | ||
| Pumping speed (Note 4) | N2 | 3200 L/s | |
| Ar | 3100 L/s | ||
| Compression ratio | N2 | 1×109 or more | |
| He | 1×105 | ||
| H2 | 6×103 | ||
| Rated speed | 27600 rpm | ||
| Start-up time | 16 minutes or less | ||
| Mounting position | In any desired direction | ||
| Vibration level (by Shimadzu's method) | 0.01 μm or less(single amplitude) | ||
| Recommended flow rate of purge gas (Note 2) | 30 mL/min | ||
| Recommended pumping speed of backing pump in case of gas purge | 1500 L/min or more | ||
| Environment temperature | 10 ~ 40 ℃ | ||
| Admissible ambient magnetic field | Radial direction | 3 mT | |
| Axial direction | 15 mT | ||
| Water | Flow Rate | 3 ~ 4 L/min | |
| Pressure | 0.2 ~ 0.4 MPa | ||
| Temperature | 19 ~ 30 degrees C.(above dew point) | ||
| Display | LED | POWER / STATUS / ROTATION / NET | |
| Communication | Contact | REMOTE(D-sub 37 pin male, Screw lock size M2.6) | |
| Input:START / STOP / RESET / LOW SPEED | |||
| Output:ROTATION / ACC. / BRAKE / NORMAL / REMOTE / ALARM / WARNING / CONNECTION | |||
| Serial | RS-232C / RS-485(D-sub 9 pin male, Screw lock size M2.6) | ||
| Speed variation | Speed is variable between 25 % and 100 % of the rated speed. (set as 0.1 %) | ||
| Input electric power | Voltage | Single phase AC 200 ~ 240 V(50 / 60 Hz) | |
| Maximum power | 1.2 kVA | ||
| Mass | 94 kg | ||
- (Note 1)Do not use corrosive gas because there is possibility that the pump breaks down by the product material.
- (Note 2)mL/min: volume flow rate at 0 degrees C., 1 atm (Compatible with SCCM).
- (Note 3)The maximum allowable flow rate depends on the gas type, the APC valve temperature, the gas purge flow rate and outlet pressure.
- (Note 4)Without a protective net. Pumping speed for N2 is 2900 L/s with a protective net.
This page may contain references to products that are not available in your country. Please contact us to check the availability of these products in your country.