Standard Specifications of Pump
| Turbo molecular pump | TMP-403LM | TMP-403LMC | TMP-403M | TMP-403MC | |
|---|---|---|---|---|---|
| Inlet flange | VG150 / ICF203 / ISO160B / ISO160C | ||||
| Outlet flange | KF25 | ||||
| Cooling method | Water | Air | |||
| Ultimate pressure (after baking) (Note 1) | 10-8 Pa order | 10-7 Pa order | 10-8 Pa order | 10-7 Pa order | |
| Maximum allowable inlet pressure (N2 continuous exhaust) | 200 Pa | 1.3 Pa | |||
| Maximum Outlet Pressure During Continuous Operation | 400 Pa | 40 Pa | |||
| Pumping speed (Note 2) | N2 | 420 L/s | |||
| He | 360 L/s | ||||
| H2 | 340 L/s | ||||
| Compression ratio | N2 | 1×109 | |||
| He | 8×104 | ||||
| H2 | 1×104 | ||||
| Rated speed | 45000 rpm | ||||
| Start-up time | 5minutes or less | ||||
| Mounting position | In any desired direction | ||||
| Pump Inlet Baking Temperature | 120 ℃ or less | ||||
| Vibration level (by Shimadzu's method) | 0.012 μm or less(single amplitude) | ||||
| Recommended flow rate of purge gas | 20 ~ 30 mL/min (Note 3) | ||||
| Recommended pumping speed of backing pump in case of gas purge | 200 L/min or more | ||||
| Environment temperature | 0 ~ 40 ℃ | ||||
| Admissible ambient magnetic field | Radial direction | 3 mT | |||
| Axial direction | 15 mT | ||||
| Water | Flow Rate | 1 ~ 3 L/min | ― | ||
| Pressure | 0.2 ~ 0.5 MPa | ||||
| Temperature | 5 ~ 30 ℃ | ||||
| Mass | 14 kg | ||||
- (Note 1)The values are for when a metal gasket is used on the inlet flange. When an O-ring gasket is used, the achieved pressure is on the order of 10^-7 Pa.
- (Note 2)Without a protective net. Pumping speed for N2 is 400 L/s with a protective net.
- (Note 3)mL/min: volume flow rate at 0 degrees C., 1 atm (Compatible with SCCM).
Standard Specifications of Power Supply Unit
| Power Supply | EI-S04M | |
|---|---|---|
| Battery | Unnecesssary | |
| Combination Compatibility | Compatible combination of pump, power supply unit, magnetic bearing cable and fan cable are available. | |
| Speed variation | Speed is variable between 25 % and 100 % of the rated speed. | |
| Communication | Contact | REMOTE(MR-34) |
| Input:START / STOP / RESET / LOW SPEED | ||
| Output:ROTATION / ACC. / BRAKE / NORMAL / REMOTE / ALARM / WARNING | ||
| Serial | Front panel:RS-232C(D-sub 9 pin male, Screw lock size M2.6) | |
| Rear panel:RS-485(D-sub 9 pin male, Screw lock size M2.6) | ||
| Permissible Ambient Conditions | Temperature | During Operation 0 ~ 45 ℃(No condensation is necessary condition. ) |
| Humidity | 40 ~ 80 %RH | |
| Input electric power | Voltage | Single phase AC 100 ~ 120 V ± 10 %(50/60 Hz ± 2Hz) |
| Single phase AC 200 ~ 240 V ± 10 %(50/60 Hz ± 2Hz) | ||
| Maximum power | 550 VA | |
| Mass | 8 kg | |