User Benefits Application News Air samples were collected using a gas sampling pump and Tenax TA sampling tube at a flow rate of 500 mL/min for 5 L. The sampling tube was then analyzed under the following analytical conditions. Sample Preparation Table 1 Analysis Conditions of Nexis GC-2060 To effectively control indoor environmental pollution, the Chinese government issued the mandatory national standard GB 50325-2020 'Code for Indoor Environmental Pollution Control of Civil Building Engineering', which uses Thermal Desorption-Gas Chromatography (TD-GC) for analysis and came into force on August 1, 2020. In this work, a method for the determination of 16 VOCs and TVOC *1 in indoor air was established using the thermal desorption/extraction injection mode of the Nexis GC-2060 gas chromatograph with MMI (Fig.1). The MMI inlet supports not only conventional split/splitless injection, but also multiple injection modes including large-volume injection (LVI), programmed temperature vaporization (PTV), thermal desorption and extraction. In the thermal desorption/extraction mode, operators only need to place the TD sampling tube into the inlet to complete thermal desorption injection. No dedicated thermal desorption device is required, resulting in cost and space saving. Introduction Using the thermal desorption/extraction injection mode of the Nexis GC-2060 MMI (multi-mode injection unit), the sampling tube is directly installed in the inlet to complete the thermal desorption injection process. All compounds meet the testing requirements of the standard GB 50325-2020 'Code for Indoor Environmental Pollution Control of Civil Building Engineering', with good sensitivity and stability. Determination of 16 Volatile Organic Compounds and TVOC in Indoor Air by Thermal Desorption-Gas Chromatography Nexis GC-2060 Gas Chromatography Chen Shen 1 ,Tianming Yao 1 1 Shimadzu (China) Co., Ltd Fig. 1 Thermal Desorption Mode for MMI Analysis Conditions The analytical condition for GC is shown in Table 1. Figure 2 shows the parameter settings for the TD injection mode in the software. Instrument: Nexis GC-2060 Column: RTX-1, 50 m×0.25 mm×0.25 μm MMI Injection Mode: Thermal Desorption/Extraction (TD/TE) Injection Mode: Split (10:1) Injection Temp.: 35 °C→900 °C/min→260 °C Carrier: He Flow Control Mode: Linear Velocity (22 cm/sec) Column Oven Temp.: 50 °C (10 min)→6 °C/min→260 °C (5 min) Detector: FID Detector Temp.: 280°C Detector Gas: Make up (He) 30 mL/min H 2 40 mL/min Air 170 mL/min Fig. 2 Parameter Settings for the TD Injection Mode Chromatograms and Calibration Curves Sixteen VOC standard compounds were weighed and dissolved in methanol to prepare mixed standard solutions with concentrations of 50, 100, 400, 800, 1200, 2000 μg/mL. The chromatogram of the 16 VOCs is shown in Figure 3. Under a nitrogen flow rate of 100 mL/min, 1 μL of each standard solution was injected into a Tenax TA sampling tubes and then the tubes were purged for 3 minutes, resulting in component masses of 50, 100, 400, 800, 1200, 2000 ng standard tubes. Perform the analysis according to the conditions in Table 1 and make external calibration curves. The linearity is shown in Figure 4 and Table 2. *1 VOCs: Volatile Organic Compounds TVOC: Total Volatile Organic Compounds
Application News 12.5 15.0 17.5 20.0 22.5 25.0 27.5 30.0 32.5 35.0 37.5 40.0 42.5 45.0 min -5000 0 5000 10000 15000 20000 25000 30000 35000 40000 uV FID1 Fig. 4 Representative Calibration Curves for VOCs No. Compounds RT(min) R 1 n-Hexane 12.77 0.9992 2 Benzene 14.341 0.9987 3 Trichloroethylene 15.622 0.9985 4 Toluene 18.446 0.9995 5 Octene 19.814 0.9995 6 Butyl acetate 20.236 0.9992 7 Ethylbenzene 22.678 0.9998 8 p-Xylene 23.068 * 0.9996 9 m-Xylene 10 o-Xylene 23.833 0.9996 11 Styrene 24.061 0.9998 12 n-Nonane 24.728 0.9997 13 2-Ethylhexanol 29.125 0.9998 14 n-Undecane 31.838 0.9998 15 n-Tetradecane 39.669 0.9994 16 n-Hexadecane 43.968 0.999 Table 2 Compound Information and Linearity of VOCs Fig. 3 Chromatogram of 16 VOCs Standard Tube Inject 1 μL of 50 μg/mL standard solution into Tenax TA tubes under 100 mL/min nitrogen flow to make 50 ng per component standard tubes. Six tubes were prepared and analyzed with the repeatability results shown in Table 3. LODs were calculated as three times the baseline noise based on 50 ng standard tubes and are listed in Table 3. Repeatability and LODs No. Compounds RSD(%) LODs (µg/m 3 ) 1 n-Hexane 2.66 1.47 2 Benzene 3.47 1.16 3 Trichloroethylene 4.10 4.72 4 Toluene 4.48 0.75 5 Octene 5.07 0.64 6 Butyl acetate 5.04 0.83 7 Ethylbenzene 4.86 0.45 8 p-/m-Xylene 4.70 0.46 9 o-Xylene 4.48 0.36 10 Styrene 4.69 0.41 11 n-Nonane 5.36 0.35 12 2-Ethylhexanol 4.89 0.40 13 n-Undecane 4.60 0.25 14 n-Tetradecane 4.03 0.25 15 n-Hexadecane 7.36 0.26 Actual Sample Analysis Table 3 Repeatability and LODs of VOCs Collect 5 L indoor air sample with a Tenax-TA tube at 500 mL/min and analyze it with the developed method. The concentration (µg/m 3 ) of target compounds and TVOC in sample are listed in Table 4. The chromatogram is shown in Figure 5. Fig. 5 Chromatogram of Actual Sample n-Hexane Trichloroethylene Toluene Octene Styrene 2-Ethylhexanol 0 500 1000 1500 浓度 0 100000 200000 300000 400000 500000 600000 700000 面积 0 500 1000 1500 浓度 0 25000 50000 75000 100000 125000 150000 175000 200000 面积 0 500 1000 1500 浓度 0 250000 500000 750000 1000000 面积 0 500 1000 1500 浓度 0 250000 500000 750000 1000000 面积 0 500 1000 1500 浓度 0 100000 200000 300000 400000 500000 600000 700000 800000 900000 1000000 面积 0 500 1000 1500 浓度 0 100000 200000 300000 400000 500000 600000 700000 800000 900000 面积 *Note: p-Xylene co-elutes with m-Xylene and the integration is combined. 1 2 3 4 5 6 7 8,9 10 11 12 13 14 15 16
Application News www.shimadzu.com/an/ Shimadzu Corporation Copyright © 2026 Shimadzu Corporation and/or its affiliates. All rights reserved. For Research Use Only. Not for use in diagnostic procedures. This publication may contain references to products that are not available in your country. Please contact us to check the availability of these products in your country. The content of this publication shall not be reproduced, altered or sold for any commercial purpose without the written approval of Shimadzu. See http://www.shimadzu.com/about/trademarks/index.html for details. Third party trademarks and trade names may be used in this publication to refer to either the entities or their products/services, whether or not they are used with trademark symbol “TM” or “”. Shimadzu disclaims any proprietary interest in trademarks and trade names other than its own. The information contained herein is provided to you "as is" without warranty of any kind including without limitation warranties as to its accuracy or completeness. Shimadzu does not assume any responsibility or liability for any damage, whether direct or indirect, relating to the use of this publication. This publication is based upon the information available to Shimadzu on or before the date of publication, and subject to change without notice. Shimadzu (China) Co., Ltd www.shimadzu.com.cn First Edition: Jul. 2026 03-GC-316-EN Conclusion A method for TVOC determination in indoor air using Tenax TA sampling tubes was established by thermal desorption/ extraction injection with MMI multi-functional inlet. All 16 VOCs showed good linearity in the range of 50 ng–2000 ng (R > 0.998). RSDs of six parallel 50 ng tests were 2.66%–7.36%. At 5 L sampling volume, MDLs were 0.25–4.72 µg/m³, which meet the requirements of TVOC testing in Appendix E of GB 50325-2020. No. Compounds Conc. 1 n-Hexane 1340 2 Benzene 1.7 3 Trichloroethylene N.D.* 4 Toluene 2.6 5 Octene N.D. 6 Butyl acetate 1.4 7 Ethylbenzene 1.7 8 p-/m-Xylene 2.3 9 o-Xylene 1.6 10 Styrene 1.2 11 n-Nonane N.D. 12 2-Ethylhexanol 2.0 13 n-Undecane N.D. 14 n-Tetradecane N.D. 15 n-Hexadecane N.D. 16 Unidentified Compounds 12.3 17 TVOC* 1360 * N.D. stands for “Not Detected”. TVOC is the sum of the measured contents of 16 VOCs and the contents of the unidentified compounds. The concentration of unidentified compounds are calculated using the toluene correction factor. Table 4 Concentrations of VOCs in Actual Samples (µg/m 3 ) Nexis is a trademark of Shimadzu Corporation or its affiliated companies in Japan and/or other countries.
Please fill out the survey Related Products Some products may be updated to newer models. Nexis GC-2060 Gas Chromatograph Related Solutions Environment Air Price Inquiry Product Inquiry Technical Service / Support Inquiry Other Inquiry