Application
Note
2
2.1. Analytical Condition
Affinisep. Commercially available samples of olive oil, coconut
oil, sesame oil, rice bran oil, and corn oil were procured from the
local market and analyzed for the target PAHs.
However, optimization of the voltages for individual MRM
transitions requires separate acquisition methods for each
transition. Subsequently, each dataset must be independently
processed and the optimized parameters manually
consolidated into a single MRM method, which significantly
increases the method development time.
To overcome this, "Create MS Table" feature of LabSolutions
GCMS software was used which automatically generates a
product ion scan method through a guided, step-by-step
workflow.
The product ion scan method generated above was used in the
MRM Optimization Tool of Smart Pesticides Database Ver.2.1,
which automatically creates an acquisition batch consisting of
multiple methods. After batch acquisition, all data files were
uploaded to the MRM Optimization Tool, which automatically
processes the individual datasets, optimize the voltages for
each MRM transition, and compile them into a single MRM
method, thereby eliminating the need for manual data
evaluation and method compilation. The final acquisition
method parameters are summarized in Table 1.
GC Inlet Parameters
Injector temperature 250 °C
Analytical column:
SH-I-5Sil MS
(P/N:221-75954-30)
(30 m x 0.25 mm;
df: 0.25 µm)
Injection mode Splitless
Injection volume 1.0 µL
Flow control mode
Linear Velocity
(55.0 cm/s)
Purge flow 5.0 mL/min
Total flow 30.0 mL/min
GC Column Oven Program
Rate (°C/min) Temperature (°C) Hold Time (min)
- 90.00 1.00
30.00 130.00 0.00
10.00 320.00 3.00
MS Parameters
Ion Source temperature 230 °C
Interface temperature 250 °C
Solvent cut time 1.5 min
Detector voltage + 0.6 kV (relative to tune result)
Target
Retention time
(min)
Type ISTD Group Target transition Reference Transitions
Acenaphthylene 5.674 Target 2 152.00>150.00 152.00>126.10; 152.00>102.10
Fluorene 7.016 Target 2 165.00>163.00 165.00>115.10; 165.00>139.10
Phenanthrene 9.077 Target 2 178.00>152.10 178.00>176.10; 176.00>150.20
Anthracene 9.191 Target 3 178.00>176.10 178.00>152.20; 176.00>150.20
Phenanthrene-D10 9.025 ISTD 2 158.00>156.10 158.00>82.10; 185.00>157.20
Anthracene-D10 9.149 ISTD 3 188.00>160.10 184.10>156.05; 160.10>158.10
Fluoranthene 11.806 Target 4 202.00>200.00 101.00>88.00; 202.00>152.50
Pyrene-D10 12.266 ISTD 4 210.00>208.10 209.00>207.20; 210.00>206.10
Pyrene 12.307 Target 4 202.00>200.10 101.00>88.00; 200.00>199.00
Benz[a]anthracene (BaA) 15.185 Target 5 228.00>226.00 114.00>101.00; 226.00>224.00
Chrysene (CHR) 15.263 Target 5 228.00>226.00 226.00>223.90; 228.00>202.20
Benzo[b]fluoranthene (BbF) 17.581 Target 5 252.00>250.00 250.00>248.00; 253.00>251.00
Benzo[k]fluoranthene 17.639 Target 5 252.00>250.00 250.00>248.10; 252.00>249.00
Benzo[a]pyrene (BaP) 18.228 Target 5 252.00>250.00 126.00>113.10; 250.00>247.30
Benzo[a]pyrene - D12 18.179 ISTD 5 261.00>257.00 116.00>102.10
Perylene-D12 18.348 ISTD 6 264.00>260.10 132.10>118.10; 132.10>130.10
Indeno[1,2,3-cd]pyrene 20.327 Target 6 276.00>274.10 138.00>125.10; 138.00>136.10
Dibenz[a,h]anthracene 20.395 Target 6 278.00>276.10 279.00>277.10; 278.00>275.00
Benzo[ghi]perylene 20.759 Target 6 276.00>274.00 138.00>125.00; 274.00>272.00
Table. 1 Analytical conditions
Fig. 2 GCMS-TQ 8050 NX
Preparation of instrument and analysis.
A scan was performed using a standard solution containing the
target PAHs and isotopically labelled internal standards. The
NIST Mass Spectral Library was used to identify the peaks for
target analytes and internal standards. The retention time and
precursor ion for each compound were established. This can be
used for creating a multiple reaction monitoring (MRM)
method.
2.2. Standard and Sample Preparation
Preparation of calibration standards.
A seven-level calibration standard was prepared using matrix-
matched solvent derived from corn oil, covering a
concentration range of 0.5-100 µg/L for the native PAHs, with
the deuterated internal standards (ISTD) maintained at a
constant concentration of 25 µg/L.
Preparation of oil samples and spiked samples.
1 mL aliquot of each edible oil sample (olive, coconut, sesame,
rice bran, and corn oil) was transferred to a suitable vial and
spiked with 50 µL of ISTD (500 µg/L).
The quantification of PAHs was carried out
using a Shimadzu GCMS-TQ8050 NX (Fig. 2),
equipped with the HTA 2800T “all-in-one GC
autosampler”. This autosampler can perform
liquid injections, headspace analysis as well
as Solid - Phase Microextraction (SPME).