Applying Non-Targeted Analysis to Food Safety
Pesticide Screening
Alan Barnes
1
, Emily Armitage
1
, Sascha Rexroth
3
, Neil Loftus
1
, Steve Williams
2
1 Shimadzu Corporation, Manchester, United Kingdom; 2 SGS Cambridge Limited, Cambridge,
United Kingdom; 3 Shimadzu Europa GmbH, Duisburg, Germany
High Performance Liquid Chromatograph Mass Spectrometer
Food commodities were measured by LC/QTOF mass spectrometry using a modified Pretreatment Optimised Injection Sequence
(POISe) to enhance retention and peak shape of polar analytes.
Non-targeted analysis (NTA) software, LabSolutions Insight Profiler, was applied to perform feature detection, data alignment,
and compound identification and successfully detected 280 test compounds at MRL concentration of 10 µg/kg.
Additional compounds outside the scope of routine analysis were identified using in-house and third-party repositories.
Introduction
Residual pesticide analysis is central to food safety, yet
conventional targeted quantitation with LC/TQ or GC/MS is
inherently limited to predefined analyte panels. Although these
platforms provide high sensitivity and quantitative precision,
expanding coverage requires labour-intensive method
development, optimisation, and validation. To address the
growing chemical complexity of agrochemical residues,
workflows are increasingly adopting high-resolution mass
spectrometry (HRMS) based non-targeted acquisition strategies.
These strategies enable simultaneous targeted quantitation and
large-scale suspect screening within a single experiment.
In this study, a unified LC/QTOF acquisition workflow was
developed to support both targeted quantitation and non-
targeted screening at scale. Leveraging large, curated screening
databases and advanced data processing with the LabSolutions
Insight Profiler software application, the method facilitates
high-confidence identification of both known and previously
unmonitored pesticide residues. This integrated strategy
significantly enhances analytical coverage while maintaining
data integrity, offering a scalable solution for next-generation
multi-residue analysis in complex food matrices.
To test the capability of non-targeted analysis (NTA), a matrix
sample (olive oil) was spiked at a concentration of 10 µg/kg, a
concentration typically defined as the maximum reporting limit
(MRL) for most pesticides. Analysis was then extended to
unknown samples to determine if additional compounds could
be identified outside the scope of routine analysis.
Materials and Methods
Analysis included a selection of unknown test samples of
various commodities previously quantified by LC/TQ. Quality
Control (QC) samples spiked with a panel of compounds were
also included. Samples were prepared using a standard
QuEChERS method.
To enhance peak shape and retention of polar analytes, sample
injection used a modified POISe method in which 1 µL of
sample was mixed with 10 µL water, repeated four times,
totalling a 4 µL sample injection. Analytical conditions were
maintained the same as the LC/TQ analysis where applicable;
LC/QTOF data acquisition utilised a non-targeted data
independent acquisition (DIA) method in positive ion mode
(Table 1).
Data Processing
Data were processed using LabSolutions Insight Profiler
software which performed feature detection, alignment and
screening using large scale suspect lists for compound
identification. Pre-defined processing parameters were used
that included non-targeted feature detection, alignment,
identification and batch processing (Fig. 2). Compounds were
identified using a curated Shimadzu pesticide database
containing over 500 compounds. Third-party libraries were also
used, including the Food Safety Mass Spectral Library
(Wageningen University)
1
.
Fig. 1 Extracted ion chromatograms of 280 pesticides spiked into olive oil at 10 µg/kg.
These were successfully found using non-targeted feature detection. The concentration broadly corresponds to the EU maximum reporting limit (MRL) in
many food commodities therefore demonstrating the application was fit for purpose for a broad panel of compounds at meaningful concentration levels.
RT (min) 1 2 3 4 5 6 7 8 9 10 11