Language
🇲🇽 Español
🇺🇸 English
0

AUTOGRAGH AGS-V Series - Specs

NUEVO

Tabletop Precision Universal Testing Machines

Especificaciones

 

Nombre del modelo AGS-1NVD/
2NVD/5NVD
AGS-10NVD/
20NVD/50NVD
AGS-
100NVD
AGS-
500NVD
AGS-
1kNVD
AGS-
5kNVD
AGS-
10kNVD
Capacidad de carga 1N/2N/5N 10N/20N/50N 100N 500N 1 kN 5 kN 10 kN
Método de carga Direct high-precision constant-speed strain control method using precision ball screw drive with no backlash
Test
Force
Exactitud High-Precision Type 1/500 Within ±0.5 % of the displayed test force (within the range of 1/1 to 1/500 of the rated load cell capacity)
JIS B7721 class 0.5, EN 10002-2 grade 0.5, ISO 7500-1 class 0.5, BS 1610 class 0.5, complies with ASTM E4 *1)
Wide-Range Accuracy Type 1/1000 Within ±1 % of the displayed test force (within the range of 1/1 to 1/1000 of the rated load cell capacity)
JIS B7721 class 1, EN 10002-2 grade 1, ISO 7500-1 class 1, BS 1610 class 1, complies with ASTM E4*1)
Standard Accuracy Type 1/500 Within ±1 % of the displayed test force (within the range of 1/1 to 1/500 of the rated load cell capacity)
JIS B7721 class 1, EN 10002-2 grade 1, ISO 7500-1 class 1, BS 1610 class 1, complies with ASTM E4*1)
Calibración de la fuerza de prueba Select from automatic test force calibration, tension, or tension and compression.
Crosshead Speed Range 0.0005 to 1,500 mm/min (can be set as desired, not in steps)
Velocidad máxima de retorno 1,650 mm/min
Crosshead Speed Accuracy *2) ±0.1 %
Velocidad de la cruceta y fuerza de prueba permitida Up to the load cell capacity for all speeds
Distance between Crosshead and Table (Tensile stroke) *3) 1,180 mm
(945 mm, SCG)
1,180 mm
(920 mm, SCG)
1,180 mm
(915 mm, SCG)
1,180 mm
(910 mm, SCG)
1,180 mm
(760 mm, MWG)
1,180 mm
(740 mm, MWG)
Ancho de prueba eficaz 425 mm
Detección de posición de la cruceta Measurement Method Codificador óptico
Display Method Pantalla digital (resolución de la pantalla: 0.001 mm)
Positional Accuracy Within ±0.1 % or within ±0.01 mm of indicated value, whichever is greater
Sampling Rate Maximum 0.2 msec
Test Condition Files 40 files
Funciones estándar
-Load cell property value automatic read in function
-Test force display function, stroke display function
-External analog output (optional, 2 channels)
-External analog input (optional, 2 channels)
-External digital input (optional, 4 channels)
-Automatic test force / strain control function (autotuning)
   -Test force automatic zero function
-Test force automatic calibration function
-Failure detection function, automatic return function
-Protection cover with interlock function*4)
-Load cell overload detection function
-Touch load detection function
Accesorios Load Cell 1 N/2 N/5 N 10 N/20 N/50 N 100 N 500 N 1 kN 5 kN 10 kN
Otro Power supply cable (2.4 m), rotation rod, DVD media (manual and other electronic data), Safety precautions sheet, Operation controller, Safety glasses, Safety gloves
Size,
Weight
Standard Model W657 mm×D525 mm×H1,562 mm
Approx. 80 kg (100 V), 75 kg (200 V)
250 mm Extension Model W657 mm×D525 mm×H1,812 mm
Approx. 85 kg (100 V), 80 kg (200 V)
500 mm Extension Model W657 mm×D525 mm×H2,062 mm
Approx. 90 kg (100 V), 85 kg (200 V)
Requisitos de suministro de corriente 1φ 100 to 115 V (100 V), 1φ 200 to 230 V (200 V) 50/60 Hz 900 VA (100 V), 1.2 kVA (200 V)
Power supply voltage fluctuations shall be within ±10 % of the installation value, and type D grounding (100 Ω or less) shall be provided.
Power Consumption 100 V:650 W、200 V:600 W
  • In the JIS B7721, EN 10002-2, ISO 7500-1, ASTM E4 standards, testing is recommended after installation of the testing machine.
  • The crosshead speed accuracy is calculated from the amount of crosshead movement within a prescribed time under steady state conditions of 0.5 mm/min to 500 mm/min.
  • The tensile stroke is the effective stroke when SCG (screw-type flat grips) or MWG (non-shift wedge-type grips) grips are installed.
  • There are options and jigs that cannot be used in conjunction with the protection cover. For more details, please contact us.

Descargas

Descargue el folleto más reciente.