High Sensitivity Analysis of Testosterone in Human Serum Using LCMS™-8060NX

Download

Introduction

This report introduces an example of high sensitivity analysis of testosterone in serum using LCMS8060NX, a triple quadrupole mass spectrometer. Calibration curves prepared using standard samples showed linearity in the concentration range of 0.05 - 1000 pg/mL (R2 =0.9996), with %RSD of concentration of 2.47% at the lower limit of quantification and accuracy at each calibration point ranged from 94.64 - 111.04%. Furthermore, the accuracy of serum samples spiked with testosterone (1 pg/mL) was 105%, indicating that the method used in the analysis is suitable for the actual samples.