Residual Solvent Analysis in Pharmaceuticals with N2 Carrier Gas Using Nexis GC-2060 and HS-20 NX — JP 18 and USP <467> / Water-Soluble Samples —

Download

User Benefits

- The newly designed FID delivers outstanding sensitivity, ensuring highly accurate and reliable analysis. - Achieve full compliance with JP and USP guidelines using N2 carrier gas, which is highly affordable and readily available. - Easily analyzes tert-butyl alcohol and Cyclopentyl methyl ether, the Class 2 solvents added in the ICH Q3C (R8) guideline.

Introduction

For the analysis of residual solvents in pharmaceuticals, headspace gas chromatography (HS-GC) is widely used to target Class 1 and Class 2 solvents, in accordance with USP General Chapter <467> and JP18 (Japanese Pharmacopoeia, 18th Edition). These regulatory tests demand high sensitivity and reproducibility. The newly developed FID for the Nexis GC-2060 offers superior detection sensitivity compared to conventional models, enabling highly sensitive analysis. Additionally, due to recent helium supply shortages and rising costs, there is a growing demand for alternative analysis using nitrogen (N2) as a carrier gas. This application note presents the analysis of Class 1 and Class 2 water-soluble samples using the Nexis GC-2060 with N2 carrier gas, compliant with Supplement II to the Japanese Pharmacopoeia. DMSO was used as the sample solvent.

August 3, 2026 GMT

Related Products

Some products may be updated to newer models