W-AD ex Series Analytical Balance

(Equipped Standard with Smart Automatic Doors and Touchless Sensors)

Model AP225W-AD ex AP135W-AD ex AP225WD-AD ex AP125WD-AD ex
Capacity 220g 135g 220g / 102g 120g / 52g
 Minimum Display  0.01mg 0.1mg / 0.01mg
Calibration Weight Built-in (*1)
External Calibration Weight Range for Span Calibration (Recommended weight value)
95 to 220.00090g
(200g)
45 to 135.00090g
(100g)
95 to 220.00090g
(200g)
45 to 120.00090g
(100g)
 Repeatability (at weighing capacity) 
(*2)
0.03mg 0.02mg 0.1mg / 0.02mg
Repeatability
(for low loads) (*2, *3)
0.0065mg (Approx. 5% load)
 Minimum Weight (*2, *3) 13mg (*2)
 Linearity (*2) ±0.1mg ±0.2mg / ±0.1mg ±0.2mg / ±0.05mg
 Response Time for Trace Measurements (*4, *5) 2 sec.
 Response Time (*5) 8 sec. 2 sec. / 8 sec.
 USB Host (Type A)
 USB Device (Type B)
Recipe Compounding
 HPLC Buffer Solution Preparation
 mol Conversion Function
 Sample (Concentration) Preparation
 Inspection Support Function
 Clock-CAL
Automatic Doors
Touchless Sensors
Adjustable Windbreak Plate
Ionizer
Operational Temperature / Humidity Range 5 to 40℃ at 20 to 85% RH (*6)
 Sensitivity Stability Against Temperature Range ±2ppm / ℃ (10~30℃)
Pan Size Approx. 91 mm dia.
Body Dimensions Approx. W215 × D411 × H346 mm (power supply unit included)
Weight Approx. 9.7kg
Display OEL display (dot matrix)
Rated Electric Power Supply DC12V 0.5A
Input Power Supply For EU type approval models: AC100V - 240V 0.48A 50/60 Hz (*7)
For other models: AC100V - 240V 0.7A 50/60 Hz (*7)
Input / Output Terminal RS-232C (D-Sub 9P plug)  USB host (Type A)  USB device (Type B)  Ionizer

*1 Minimum display 0.01 mg models provide two internal weights as standard.

*2 Based on our shipping inspection conditions. 

*3 Compliant with the Japanese Pharmacopoeia, USP 41 and Ph.Eur.2.1.7. Minimum weight value is met at the Shimadzu Corporation factory using weights equal to approximately 5 % of the weighing capacity. This value is affected by the installation environment, so it is necessary to measure it in the actual environment of use.

*4 The response time for displaying 90 % of the added sample amount value in trace measurements (from 1 mg).

*5 The response time value is typical.

*6 Non-condensing.

*7 The Input Power Supply may change without notice.

W ex Series Analytical Balance

Model AP225W ex AP135W ex AP225WD ex AP125WD ex
Capacity 220g 135g 120g / 52g 220g / 102g
Minimum Display 0.01mg 0.1mg / 0.01mg
Calibration Weight Built-in (*1)
External Calibration Weight Range for Span Calibration (Recommended weight value)
95 to 220.00090g
(200g)
45 to 135.00090g
(100g)
45 to 120.00090g
(100g)
95 to 220.00090g
(200g)
 Repeatability (at weighing capacity)
(*2)
0.03mg 0.02mg 0.1mg / 0.02mg
Repeatability
(for low loads) (*2, *3)
0.0065mg (approx. 5 % load)
 Minimum Weight (*2, *3) 13mg
Linearity (*2) ±0.1mg ±0.2mg / ±0.1mg ±0.2mg / ±0.05mg
 Response Time for Trace Measurements (*4, *5) 2 sec.
 Response Time (*5) 8 sec. 2 sec. / 8 sec.
 USB Host (Type A)
USB Device (Type B)
Recipe Compounding
HPLC Buffer Solution Preparation
mol Conversion Function
Sample (Concentration) Preparation
Inspection Support Function
Clock-CAL
Internal Windbreak Plate
Ionizer (optional)
Operational Temperature / Humidity Range 5 to 40℃ at 20 to 85% RH (*6)
Sensitivity Stability Against Temperature Range ±2ppm/℃ (10~30℃)
Pan Size Approx. 91 mm dia.
Body Dimensions Approx. W213 × D411 × H345 mm (power supply unit included)
Weight Approx. 7.9 kg
Display OEL display (dot matrix)
Rated Electric Power Supply DC12V 0.3A
Input Power Supply For EU type approval models: AC100V - 240V 0.32 - 0.19 A 50/60 Hz (*7)
For other models: AC100 V - 240 V 0.7 A 50/60 Hz (*7)
Input / Output Terminal RS-232C (D-Sub 9P plug)  USB host (Type A)  USB device (Type B)  Ionizer

*1 Minimum display 0.01 mg models provide two internal weights as standard.

*2 Based on our shipping inspection conditions. 

*3 Compliant with the Japanese Pharmacopoeia, USP 41 and Ph.Eur.2.1.7. Minimum weight value is met at the Shimadzu Corporation factory using weights equal to approximately 5 % of the weighing capacity. This value is affected by the installation environment, so it is necessary to measure it in the actual environment of use.

*4 The response time for displaying 90 % of the added sample amount value in trace measurements (from 1 mg).

*5 The response time value is typical.

*6 Non-condensing.

*7 The Input Power Supply may change without notice.

W-AD Series Analytical Balance

(Equipped Standard with Smart Automatic Doors and Touchless Sensors)

Model AP225W-AD AP135W-AD AP225WD-AD AP125WD-AD AP324W-AD AP224W-AD
Capacity 220g 135g 220g / 102g 120g / 52g 320g 220g
 Minimum Display  0.01mg 0.1mg / 0.01mg 0.1mg
Calibration Weight Built-in (*1)
External Calibration Weight Range for Span Calibration (Recommended weight value)
95 to 220.00090g
(200g)
45 to 135.00090g
(100g)
95 to 220.00090g
(200g)
45 to 120.00090g
(100g)
95 to 320.0090g
(300g)
95 to 220.0090g 
(200g)
 Repeatability (at weighing capacity) 
(*2)
0.015mg (up  to 20g)
0.03mg (up to 100g)
0.05mg (up to 220g)
0.05mg 0.1mg / 0.05mg 0.1mg / 0.02mg 0.15mg 0.1mg
Repeatability
(for low loads) (*2, *3)
0.01mg
(approx. 5% load)
0.1mg
(approx. 5% load)
 Minimum Weight (*2, *3) 20mg 200mg
 Linearity (*2) ±0.1mg ±0.2mg / ±0.1mg ±0.2mg / ±0.05mg ±0.3mg ±0.2mg
 Response Time for Trace Measurements (*4, *5) 2 sec.
 Response Time (*5) 8 sec. 2 sec. / 8 sec. 2 sec.
 USB Host (Type A)
 USB Device (Type B)
Recipe Compounding
 HPLC Buffer Solution Preparation
 mol Conversion Function
 Sample (Concentration) Preparation
 Inspection Support Function
 Clock-CAL
Automatic Doors
Touchless Sensors
Adjustable Windbreak Plate -
Ionizer
Operational Temperature / Humidity Range 5 to 40℃ at 20 to 85% RH (*6)
 Sensitivity Stability Against Temperature Range ±2ppm / ℃ (10~30℃)
Pan Size Approx. 91 mm dia.
Body Dimensions Approx. W215 × D411 × H346 mm (power supply unit included) Approx. W215 × D367 × H346 mm
Weight Approx. 9.7kg Approx. 8.6kg
Display OEL display (dot matrix)
Rated Electric Power Supply DC12V 0.5A
Input Power Supply For EU type approval models: AC100V - 240V 0.48A 50/60 Hz (*7)
For other models: AC100V - 240V 0.7A 50/60 Hz (*7)
Input / Output Terminal RS-232C (D-Sub 9P plug)  USB host (Type A)  USB device (Type B)  Ionizer

*1 Minimum display 0.01 mg models provide two internal weights as standard.

*2 Based on our shipping inspection conditions. 

*3 Compliant with the Japanese Pharmacopoeia, USP 41 and Ph.Eur.2.1.7. Minimum weight value is met at the Shimadzu Corporation factory using weights equal to approximately 5 % of the weighing capacity. This value is affected by the installation environment, so it is necessary to measure it in the actual environment of use.

*4 The response time for displaying 90 % of the added sample amount value in trace measurements (from 1 mg).

*5 The response time value is typical.

*6 Non-condensing.

*7 The Input Power Supply may change without notice.

W Series Analytical Balance

Model AP225W AP135W AP125WD AP225WD AP124W AP224W AP324W
Capacity 220g 135g 120g / 52g 220g / 102g 120g 220g 320g
Minimum Display 0.01mg 0.1mg / 0.01mg 0.1mg
Calibration Weight Built-in (*1)
External Calibration Weight Range for Span Calibration (Recommended weight value)
95 to 220.00090g
(200g)
45 to 135.00090g
(100g)
45 to 120.00090g
(100g)
95 to 220.00090g
(200g)
45 to 120.0090g
(100g)
95 to 220.0090g
(200g)
95 to 320.0090g
(300g)
 Repeatability (at weighing capacity)
(*2)
0.015mg (up  to 20g)
0.03mg (up to 100g)
0.05mg (up to 220g)
0.05mg 0.1mg / 0.02mg 0.1mg / 0.05mg 0.1mg 0.15mg
Repeatability
(for low loads) (*2, *3)
0.01mg (approx. 5% load) 0.1mg
(approx. 5% load)
 Minimum Weight (*2, *3) 20mg 200mg
Linearity (*2) ±0.1mg ±0.2mg / ±0.05mg ±0.2mg / ±0.1mg ±0.2mg ±0.3mg
 Response Time for Trace Measurements (*4, *5) 2 sec.
 Response Time (*5) 8 sec. 2 sec. / 8 sec. 2 sec.
 USB Host (Type A)
USB Device (Type B)
Recipe Compounding
HPLC Buffer Solution Preparation
mol Conversion Function
Sample (Concentration) Preparation
Inspection Support Function
Clock-CAL
Internal Windbreak Plate ○ (optional)
Ionizer (optional)
Operational Temperature / Humidity Range 5 to 40℃ at 20 to 85% RH (*6)
Sensitivity Stability Against Temperature Range ±2ppm/℃ (10~30℃)
Pan Size Approx. 91 mm dia.
Body Dimensions Approx. W213 × D411 × H345 mm (power supply unit included) Approx. W213 × D367 × H345 mm
Weight Approx. 7.9 kg Approx. 7.0 kg
Display OEL display (dot matrix)
Rated Electric Power Supply DC12V 0.3A
Input Power Supply For EU type approval models: AC100V - 240V 0.32 - 0.19 A 50/60 Hz (*7)
For other models: AC100 V - 240 V 0.7 A 50/60 Hz (*7)
Input / Output Terminal RS-232C (D-Sub 9P plug)  USB host (Type A)  USB device (Type B)  Ionizer

*1 Minimum display 0.01 mg models provide two internal weights as standard.

*2 Based on our shipping inspection conditions. 

*3 Compliant with the Japanese Pharmacopoeia, USP 41 and Ph.Eur.2.1.7. Minimum weight value is met at the Shimadzu Corporation factory using weights equal to approximately 5 % of the weighing capacity. This value is affected by the installation environment, so it is necessary to measure it in the actual environment of use.

*4 The response time for displaying 90 % of the added sample amount value in trace measurements (from 1 mg).

*5 The response time value is typical.

*6 Non-condensing.

*7 The Input Power Supply may change without notice.

X Series / Y Series Analytical Balance

Series X Series Y Series
Model AP324X AP224X AP124X AP324Y AP224Y AP124Y
Capacity 320g 220g 120g 320g 220g 120g
Minimum Display 0.1mg
Calibration Weight Built-in None
External Calibration Weight Range for Span Calibration (Recommended weight value)
95 to 320.0090g
(300g)
95 to 220.0090g
(200g)
45 to 120.0090g
(100g)
95 to 320.0090g
(300g)
95 to 220.0090g
(200g)
45 to 120.0090g
(100g)
Repeatability (at weighing capacity) 0.15mg 0.1mg 0.15mg 0.1mg
Repeatability
(for low loads) (*1)
0.1mg (approx. 5% load) 
Minimum Weight (*1) 200mg
Linearity ±0.3mg ±0.2mg ±0.3mg ±0.2mg
 Response Time for Trace Measurements (*2, *3) 2 sec.
 Response Time (*3) 2 sec.
 USB Host (Type A)
USB Device (Type B)
Recipe Compounding
HPLC Buffer Solution Preparation
mol Conversion Function
Sample (Concentration) Preparation
Inspection Support Function
Clock-CAL
Ionizer
Operational Temperature / Humidity Range 5~40℃ 20 to 85% (*4)
Sensitivity Stability Against Temperature Range ±2ppm/℃ (10 to 30℃)
Pan Size Approx. 91 mm dia.
Body Dimensions Approx. W213 × D367 × H345 mm
Weight Approx. 7.0 kg Approx. 6.5 kg
Display OEL display (dot matrix)
Rated Electric Power Supply DC12V 0.3A
Power Supply Input For EU type approval models: AC100V - 240V 0.32 - 0.19 A 50/60 Hz (*5)
For other models: AC100 V - 240 V 0.7 A 50/60 Hz (*5)
Input / Output Terminal RS-232C (D-Sub 9P plug)  USB device (Type B)  Ionizer RS-232C (D-Sub 9P plug)  USB device (Type B) 

*1 Compliant with the Japanese Pharmacopoeia, USP 41 and Ph.Eur.2.1.7. Minimum weight value is met at the Shimadzu Corporation factory using weights equal to approximately 5 % of the weighing capacity. This value is affected by the installation environment, so it is necessary to measure it in the actual environment of use.

*2 The response time for displaying 90 % of the added sample amount value in trace measurements (from 1 mg).

*3 The response time value is typical.

*4 Non-condensing.

*5 The Input Power Supply may change without notice.

{"title":"Specification Sheet Downloads","description":"Download the latest specification sheets.","source":"product","key":3525,"max":30,"filter_types":["spec_sheet"],"link_title":"View other Downloads","link_url":""} {"title":"Downloads","description":"Download the latest brochure.","source":"product","key":3525,"max":30,"filter_types":["brochures"],"link_title":"View other Downloads","link_url":"","pdf_links":[]}