AP Series - The recipe preparation mode
1. How to register recipe preparation
-
[Key name]

-
[Side of the unit]

Selecting an Item, Entering Numbers and Text
Use the arrow keys to select an item. At the cursor, use the Up and Down keys to change a number or character. You can also use USB keyboard to select an item and change a number or character
-
1.Set the Recipe preparation
- Press [
MENU ] while in the weight measurement mode, and then press [
LEFT ]. - Select [
Select Mode ], and then press [
OK ]. - Select [
Recipe preparation ], and then press [
OK ].
- Press [
-

-
2.Select the recipe
- Select any of the recipes from [RECIPE1-5] (or the recipe name), and then press [
OK ].
- Select any of the recipes from [RECIPE1-5] (or the recipe name), and then press [
-

-
3.Set the Recipe name
- Select [Recipe name], and then press [
OK ]. - Enter the recipe name, and then press [
OK ].
- Select [Recipe name], and then press [
-

-
4.Set taring
- Select [Taring], and then press [
OK ]. - Select [Auto] or [Manual], and then press [
OK ].
Taring Setting
If [Automatic] is selected, taring will be carried out automatically when the next sample is placed in the pan after the recipe preparation when sample weighing has been completed. If [Manual] is selected, taring needs to be carried out manually for each sample preparation. - Select [Taring], and then press [
-

-
5.Register the ingredients
- Select any of the ingredients from [Component 1-10] (or ingredient name), and then press [
OK ], and select [Setting] then press [
OK ]. - Enter the ingredient name within 20 characters, and then press [
OK ]. - Select the unit, and then press [
OK ]. - Enter the weight, and then press [
OK ]. - Enter the permissible error value, and then press [
OK ].
Enable Weighing Units in Advance
By default, only g (grams) is enabled as a weighing unit. On the weighing display, press and hold [
Up ] or at least 3 seconds to set up other units.
Enable other units in advance to make them available for selection.
If the unit of each component is not unified.
Units of the total value is “g (gram)”.
[Permissible Error Value] is...
The [Permissible Error Value] entered in Step 5 represents the value for which an OK judgment is permitted with regard to the range of ingredient weight for each ingredient. - Select any of the ingredients from [Component 1-10] (or ingredient name), and then press [
-

-
6.Repeat Step 5. for the required number of samples
-
7.Return to the weight measurement mode
- Press [
POWER ].
Returning to Standard Weighing Mode
From Recipe Formulation Mode, press [
Down ] to return to Standard Weighing Mode.
Press [
Down ] again to resume the most recently used recipe. - Press [
-

2. How to use recipe preparation
1.Set the Recipe preparation
- Press [
MENU ] while in the weight measurement mode, and then press [
LEFT ]. - Select [
Select Mode ], and then press [
OK ]. - Select [
Recipe preparation ], and then press [
OK ].
2.Select the recipe
Select any of the recipes from [RECIPE1-5] (or the recipe name), and then press [
OK ].
-
3.Measure the first sample
- Place the container on the pan.
- Press [
O/T ]. - Weigh the sample in accordance with the ingredient name, weight and permissible error value displayed on the screen.
- Press [
OK ] when
(Stability Mark) is displayed.
-

-
4.Formulate the sample
- Press [
O/T ]. - Weigh the sample in accordance with the ingredient name, weight and permissible error value displayed on the screen.
- Press [
OK ] when
(Stability Mark) is displayed.
Taring Setting
It is not necessary to press [
O/T ] as instructed in 1 if [Taring] has been set at [First Time Only]. - Press [
-

5.Repeat Step 4. for the required number of samples
-
6.End Sample preparation
The total weight will be displayed when Sample preparations for all ingredients have been completed.
Read the display, and then press [
OK ].
Recipe preparation data output
Once a recipe has been prepared, the output process will always be started and [Printing] will be displayed to make sure that the weight record of the prepared sample is saved. -

7.Cancel the hold function
Press [
OK ].