ShimNex Ghost Trap - Features

Mobile phase cleaner for HPLC

Effectively Removes Ghost Peaks

  • Installing ShimNex Ghost Trap between the gradient mixer and the autosampler enables the removal of impurities contained in the mobile phase as well as contaminants originating from the piping and gradient mixer. The chromatogram shown below was obtained by performing gradient analysis using a phosphate buffer solution without sample injection. When no mobile phase cleaner was used, impurities in the mobile phase appear as peaks in the chromatogram. In contrast, when ShimNex Ghost Trap was used, these impurities were efficiently removed, and the ghost peaks disappeared.
  • ShimNex Ghost Trap

Comparison of Ghost Trap Performance

  • Column ShimNex CS C18 (5 μm, 4.6 × 150 mm)*
    Flow Rate 1 mL/min
    Temperature 30 ℃
    Detection PDA, 220 nm
    Mobile Phase A: 50 mmol/L NaH2PO4-Na2HPO4 (pH 6.8)
    B: Acetonitrile

    * ShimNex CS C18 is sold by Shimadzu (Shanghai) Global Laboratory Consumables Co., Ltd. 

  • chromatogram

Elution Gradient Conditions

Time (Min) 0 20 20.5 30
A (%) 95 5 95 95
B (%) 5 95 5 5

Lineup Compatible with Both HPLC and UHPLC

Two ShimNex Ghost Trap models are available: the cartridge type and the packed type for use with UHPLC systems. Select the appropriate one for your application.

PN Description Dimensions Pressure resistance
380-01259-92 Cartridge 2 pc 7.6 mmI.D.×30 mm 40 MPa
380-01259-94 Cartridge 1 pc
Holder 1 pc
380-01259-93 Cartridge 2 pc 4.0 mmI.D.×20 mm 40 MPa
380-01259-95 Cartridge 1 pc
Holder 1 pc
380-01259-96 Packed type 1 pc 2.1 mmI.D.×30 mm 70 MPa
380-01259-97 Packed type 1 pc 4.6 mmI.D.×50 mm 70 MPa

* Installing the product downstream of the gradient mixer or the confluence of two pumps will introduce a delay volume equivalent to the internal volume of the product.
* Product service life depends on the analytical conditions, such as the mobile phase used.
* When using an ion-pairing reagent, it may be retained in the product, potentially affecting retention times and peak shapes.
* Before connecting the analytical column, thoroughly clean the flow path with mobile phase (close to the final concentration for gradient analysis).
* Some impurities may not be removed.