Q: If acetonitrile is replaced with methanol, the pressure applied to the column increases. Is there any influence on the UV detector cell?

    A:

    The column is the part in HPLC channels that is subjected to the greatest pressure. Beyond the column outlet, there is no resistance and so hardly any pressure is applied. A methanol-based mobile phase can be used with a standard detector cell without any problems.