User Benefits
Application
News
Instrument: TOC-L total organic carbon analyzer
Oxidation Method: 680 °C combustion catalytic oxidation
Catalyst: High sensitivity catalyst
Measurement Items: NPOC (TOC by acidification and sparging)
TOC (TOC by the TC-IC method)
Calibration Curves: NPOC and TC: 2-point calibration curves based on
potassium hydrogen phthalate aqueous solutions
at 0 and 5 mgC/L
IC: 2-point calibration curve based on sodium
bicarbonate/sodium carbonate aqueous solutions
at 0 and 5 mgC/L
Injection Volume: 200 μL
Sample: Commercial sodium dodecylbenzenesulfonate
reagent was dissolved in pure water
Samples and Analysis Conditions
Four sodium dodecylbenzenesulfonate solutions were prepared
as surfactant samples, at TOC concentrations of 0.5, 1, 2, and
3 mg/L, respectively (Table 1). TOC measurements of these
samples were conducted under the conditions indicated in
Table 2. The following two measurement methods were used.
(1) TC-IC method: TC (total carbon) and IC (inorganic carbon)
are measured separately, and then TOC is calculated as their
difference (TOC = TC – IC).
(2) NPOC method: TOC is measured as non-purgeable organic
carbon (NPOC). After acidifying the sample to a pH < 3, IC is
subsequently removed by sparging and the remaining
carbon is measured as TC, which is regarded as TOC. The
acidification and sparging steps are performed
automatically by the instrument.
Table 1 Samples
Table 2 TOC Analysis Conditions
Introduction
Surfactants are important chemicals widely used in detergents
and dispersants for pharmaceutical equipment cleaning
processes, chemical manufacturing processes, and
environmental water management. If surfactant residues
remain after such processes, they can not only affect
manufacturing processes and final product quality, but also
increase the organic load in wastewater and environmental
water. Therefore, quality control is also needed for organic
matter derived from surfactants.
Total organic carbon (TOC) analysis is effective for evaluating
the quality of surfactants. In general, the NPOC (non-purgeable
organic carbon) method is often selected for TOC measurement
in terms of sample volume and analysis time requirements. On
the other hand, because surfactants have both hydrophilic and
hydrophobic groups in their structure, they tend to foam even
at low concentrations. When such samples are measured by the
NPOC method, which involves acid addition and sparging,
foaming can cause the sample to overflow from the syringe
vessel inside the instrument into the waste line. That can result
in measurement values that are lower than the true value.
This article compares TOC measurement results for surfactants
obtained by the NPOC method and the TC-IC method using the
Shimadzu TOC-L total organic carbon analyzer, which uses the
combustion oxidation method.
TOC can be evaluated accurately using the TC-IC method, even for foaming surfactants.
The ASI-L autosampler enables automatic, continuous measurement of multiple samples, improving analytical efficiency.
TOC Evaluation of Surfactants
Arihiro Narita
Total Organic Carbon Analyzer
Sample
Added Carbon Concentration (mg/L)
in Sodium Dodecylbenzenesulfonate
1 0.5
2 1
3 2
4 3
Calibration Curves
The TOC analyzer was calibrated using 2-point calibration
curves for potassium hydrogen phthalate at 0 and 5 mgC/L and
2-point calibration curves for sodium bicarbonate/sodium
carbonate aqueous solutions at 0 and 5 mgC/L. The potassium
hydrogen phthalate calibration curve was used for NPOC and TC
measurements and the sodium bicarbonate/sodium carbonate
calibration curve was used for IC measurement.
The calibration curves are shown in Figs. 1 (a) to (c).
Fig. 1 (a) Calibration Curve Data (NPOC Measurement)
Fig. 1 (b) Calibration Curve Data (TC Measurement)
Fig. 1 (c) Calibration Curve Data (IC Measurement)