User Benefits
Application
News
Fig. 1 Analysis Workflow Using LabSolutions Insight Profiler
LabSolutions Insight Profiler
LabSolutions Insight Profiler is software for batch processing of
LC-QTOFMS data that enables non-targeted analysis through an
intuitive workflow. It readily supports easy spectral feature
detection, alignment, statistical processing, filtering, list
screening, MS/MS spectra-based library searching, principal
component analysis (PCA), displaying volcano or box plots, and
more. Fig. 1 shows the data analysis workflow.
Table 2 Analysis Conditions
Data Analysis
LabSolutions Insight Profiler was used for data analysis. Fig. 2
shows an example of the LabSolutions Insight Profiler analysis
screen. Alignment results, chromatograms, and spectra are
linked with PCA and volcano plots, facilitating the interpretation
of complex datasets. For library searches, the NIST 23 Mass
Spectral Library, which contains approximately 50,000
compounds, was used. Scores were calculated based on MS/MS
spectral matching.
Sample and Pretreatment
Six commercially available beer-type beverages, including beer,
low-malt beer, beer-like beverage, and non-alcoholic beer, were
used as samples (Table 1). Each sample was pretreated by
degassing and then 10-fold dilution with ultrapure water.
Introduction
Metabolomics, a technique that analyzes metabolites in living
organisms, is increasingly being applied in the food industry for
product development and quality control. A particularly powerful
approach is non-targeted metabolomics, which aims to analyze
not only known metabolites but also unknown compounds
without prior assumptions to reveal unexpected chemical
differences between samples. Quadrupole time-of-flight mass
spectrometry (QTOFMS), with its accurate-mass capability, is well
suited for this purpose. However, the large volume of generated
data creates challenges for feature identification and data
comparison.
This article presents a case study in which LabSolutions Insight
Profiler, a software platform for non-targeted analysis, was used
to characterize compounds in beer and beer-like beverages. Peak
detection, retention-time alignment, and library searching were
performed across multiple datasets. Multivariate analysis revealed
compound profiles that likely reflect the differences in raw
materials and manufacturing processes, demonstrating the utility
of the approach for beverage characterization.
Non-Targeted Analysis of Metabolites in Alcoholic
Beverages Using LabSolutions Insight Profiler
Yutaka Umakoshi and Tetsuo Iida
High Performance Liquid Chromatograph Mass Spectrometer
No Sample Description
1 Beer 1 Lager beer (bottom fermentation)
2 Beer 2 Ale beer (top fermentation)
3 Low-malt beer Purine free
4 Beer-like beverage Contains soy protein
5 Non-alcoholic beer 1 Made in Japan
6 Non-alcoholic beer 2 Made in Germany
Table 1 Sample Information
HPLC Conditions (Nexera X3)
Column: Reversed-phase column
Column Oven: 40 °C
Solvent A: 0.1 % Formic acid in water
Solvent B: 0.1 % Formic acid in acetonitrile
Mode: Gradient elution
Flowrate: 0.25 mL/min
Injection Volume: 3 µL
MS Conditions (LCMS-9030)
Ionization: ESI, Positive
Mode: Data dependent acquisition (DDA)
MS Event: m/z 50-1000, 0.1 sec
MS/MS Event: m/z 10-1000, 0.1 sec x5
Collision Energy: 35.0 ± 17.0 V
Nebulizing Gas Flow: 3.0 L/min
Drying Gas Flow: 10.0 L/min
Heating Gas Flow: 10.0 L/min
DL Temp.: 250 °C
Block Heater Temp.: 400 °C
Interface Temp.: 300 °C
CID Gas Pressure: 230 kPa
6.2 6.4 6.6 6.8 7.0
0.0e0
5.0e4
1.0e5
1.5e5
2.0e5
Principal Component Analysis (PCA) Volcano Plot Box Plot
Feature Detection/Alignment Library Search
MS1 data
or
DDA data
LabSolutions Insight Profiler
51.02000 65.04000
77.04000
95.05000
m/z 20 40 60 80 100
%
100.00
51.02245
65.03817
77.03813
95.04872
119.04886
0.00
%
100.00
Using an LC-QTOFMS system with LabSolutions Insight Profiler non-targeted analysis software enables comprehensive
comparative analysis of metabolites in food samples.
Principal component analysis and volcano plots can be used to detect characteristic features in each sample.
A variety of functions, including MS/MS spectral library searching, enable easy qualitative analysis of components.
Analysis Conditions
Nexera X3 UHPLC and LCMS-9030 systems were used as the
analytical instruments. The LC method included in the
LC/MS/MS Method Package for Primary Metabolites was used as
the method. The data-dependent acquisition (DDA) mode was
used to simultaneously acquire precursor m/z and MS/MS data.
Table 2 shows the analysis conditions.