Application
News
www.shimadzu.com/an/
Shimadzu Corporation
© Shimadzu Corporation, 2026
For Research Use Only. Not for use in diagnostic procedures.
This publication may contain references to products that are not available in your country. Please contact us to check the availability of these
products in your country.
The content of this publication shall not be reproduced, altered or sold for any commercial purpose without the written approval of Shimadzu.
See https://www.shimadzu.com/about/trademarks/index.html for details.
Third party trademarks and trade names may be used in this publication to refer to either the entities or their products/services, whether or not
they are used with trademark symbol “TM” or “”.
Shimadzu disclaims any proprietary interest in trademarks and trade names other than its own.
The information contained herein is provided to you "as is" without warranty of any kind including without limitation warranties as to its
accuracy or completeness. Shimadzu does not assume any responsibility or liability for any damage, whether direct or indirect, relating to the
use of this publication. This publication is based upon the information available to Shimadzu on or before the date of publication, and subject
to change without notice.
First Edition: Aug. 2026 01-01195-EN
Table 3 Analysis Conditions (Calibration Curve Method)
Element V, Cr, Mn, Co, Ni, Cu, Zn, As, Se, Mo, Cd, Sn, Hg, Tl, Pb
Analytical Line
VKα
,
CrKα
,
MnKα
,
CoKα
,
NiKα
,
CuKα
,
ZnKα
,
AsKα
,
SeKα
,
MoKα
,
CdKα
,
SnKα
,
HgLα
,
TlLα
,
PbLβ1
Analysis
Conditions
Calibration Curve Method
(Scattered X-ray Internal Standard Method)
Detector SDD Detector
X-ray Tube Rh Target
Tube Voltage
30 kV (V, Cr, Mn, Co, Ni),
50 kV (Cu, Zn, As, Se, Mo, Cd, Sn, Hg, Tl, Pb)
Tube Current Auto [µA]
Collimator 10 mmφ
Primary Filter
#5 (V), #3 (Cr, Mn, Co, Ni),
#4 (Cu, Zn, As, Se, Hg, Tl, Pb), #1 (Mo, Cd, Sn)
Atmosphere Air
Integration Time
60 s
×
4 ch
(120 s
×
4 ch for Calibration Curve Samples Only)
Dead Time Max. 30 %
Element Na-U
Analytical Line Kα, Kβ, Lα, Lβ
・・・
Analysis
Conditions
Qual-Quantitative Analysis (Thin-Film FP Method
*1
)
Detector SDD Detector
X-ray Tube Rh Target
Tube Voltage
15 kV (Na-S), (Cl-V),
50 kV (Cr-Co), (Ni-Nb, As, Pb), (Mo-Xe)
Tube Current Auto [µA]
Collimator 10 mmφ
Primary Filter None (Na-S), #2 (Cl-V), #4 (Ni-Nb, As, Pb), #1 (Mo-Xe)
Atmosphere Air
Integration
Time
60 s
×
5 ch
Dead Time Max. 30 %
Table 4 Analysis Conditions (Qual-Quantitative Analysis)
Table 2 Quantitative Analysis Results [µg/mL]
Quant.
23V 24Cr 25Mn 27Co 28Ni
Value 3σ Accuracy Value 3σ Accuracy Value 3σ Accuracy Value 3σ Accuracy Value 3σ Accuracy
Calib. 8.8 2.1 88% 10.9 1.5 109% 9.1 1.1 91% 9.5 1.3 95% 9.6 1.0 96%
FP
*1
8.7 1.8 87% 9.7 1.8 97% 8.5 1.3 85% 10.0 1.3 100% 9.0 0.8 90%
Quant.
29Cu
*2
30Zn
*3
33As 34Se 42Mo
Value 3σ Accuracy Value 3σ Accuracy Value 3σ Accuracy Value 3σ Accuracy Value 3σ Accuracy
Calib. 10.9 0.7 93% 14.3 0.6 84% 9.9 0.6 99% 10.2 0.3 102% 10.0 0.9 100%
FP
*1
11.3 0.5 97% 15.3 0.5 94% 9.4 0.3 94% 10.3 0.3 103% 9.0 0.7 90%
Quant.
48Cd 50Sn 80Hg 81Tl 82Pb
Value 3σ Accuracy Value 3σ Accuracy Value 3σ Accuracy Value 3σ Accuracy Value 3σ Accuracy
Calib. 11.8 1.7 118% 9.5 2.9 95% 10.3 0.5 103% 9.9 0.5 99% 10.0 0.8 100%
FP
*1
9.8 1.7 98% 9.0 2.8 90% 9.6 0.4 96% 9.7 0.4 97% 8.4 0.6 84%
*1 In the FP method, moisture and organic components that cannot be measured were treated as CH₂O balance (remainder) and quantified.
*2 Considering that the blood sample before standard addition contained 1.6 µg/mL of Cu, the accuracy was calculated using the following equation.
Accuracy %=
Value [ppm]−1.6
10.0
×100
*3 Considering that the blood sample before standard addition contained 5.9 µg/mL of Zn, the accuracy was calculated using the following equation.
Accuracy %=
Value [ppm]−5.9
10.0
×100
Conclusion
The analysis conditions for the calibration curve method are
shown in Table 3, and those for the qual-quantitative analysis
using the FP method are shown in Table 4.
Analysis Conditions
The FP method demonstrated high accuracy for quantitation
without calibration curves. In addition, blood samples do not
require complicated pretreatment such as deproteinization, and
the analysis time is only about 10 minutes.
*1 For a light-element matrix such as blood, the thin-film FP method was
applied and the sample thickness was set, taking into account the
greater analysis depth for heavy elements.