Application
News
Fig. 5 Analysis Screens of LabSolutions Insight Explore GCMS
Fig. 4 Simple Concentration of Aroma Components Using MonoTrap
(Left: Vapor-Phase Extraction, Right: Immersed Extraction)
Sample Preparation and Analysis Conditions
The samples used in this experiment were two types (A and B)
of commercially available soy sauce. Using MonoTrap, a
monolithic silica adsorbent, the aroma components were
concentrated by vapor-phase extraction and immersion (liquid-
phase) extraction (Fig. 4). Specifically, 6 mL of the sample and a
MonoTrap RGPS TD (GL Sciences) were introduced into a 20 mL
screw vial and concentrated under either vapor-phase or
immersion extraction conditions while stirring at 250 rpm at
60 ˚C for 30 min. After concentration, the excess moisture
adhering to the MonoTrap was wiped off with a Kimwipe, and
the MonoTrap was placed in a splitless glass insert (Xtra Inert
Liner, Splitless, P/N: 227-36701-01) for use with the MMI.
MonoTrap
Table 1 shows the instrument configuration used here, together
with the analysis conditions. A capillary tube was connected to
the column as a post-column. This made it possible to prevent
deterioration of the column liquid phase at the interface side,
even when exchanging the insert liner, without lowering the
interface temperature. Use of the Easy sTop function also made
it possible to exchange the glass insert safely and reliably by
displaying the permissible timing for the replacement and
automatically checking for leaks.
Table 1 Instrument Configuration and Analysis Conditions
Configuration
GC-MS
:
GCMS-TQ8040 RX
Injection Unit : MMI-U
Column : InertCap Pure-WAX
(30 m × 0.25 mm I.D., 0.25 μm)
(GL Sciences Inc.)
*A post-column (30 cm × 0.25 mm I.D.) was connected
with a micro union.
GC
MMI Mode : Thermal desorption/extraction introduction method
Purge Program : [Time] 0 min
[Temperature] 35 ˚C
[Split Ratio] 10
Desorb Program : [Time] 10 min
[Temperature] 250 ˚C
[Heating Rate] 1200 ˚C/min
[Split Ratio] 10
Carrier gas : He
Carrier gas control : Pressure (83.5 kPa)
Column temp. : 50 ˚C (5 min) – 10 ˚C/min – 250 ˚C (10 min)
MS
Ion source temp.
:
200 ˚C
Interface temp. : 250 ˚C
Acquisition mode : Scan/MRM (using conditions of Smart Aroma Database)
Loop Time : Scan 0.1 s, MRM 0.2 s
Data Analysis
The simultaneous scan/MRM measurement data were analyzed
using LabSolutions Insight Explore GCMS (Fig. 5). For scan data,
this software can perform peak detection by deconvolution and
a library search. The same program can also carry out targeted
analysis and quantitative analysis of MRM data.
Targeted analysis of MRM data
Non-targeted analysis of scan data
Comprehensive Search Using Scan Data
Peak detection by deconvolution and identification of the
detected compounds by a library search were carried out for the
scan data. In the mass spectrum library, the mass spectrum
library (AROMA_IC-WAX) belonging to the Smart Aroma
Database and the NIST mass spectrum library (NIST23) were
used, and a similarity score of 80 was used as the threshold.
Among the detected peaks, peaks originating from collision gas
and peaks of siloxane originating from the MonoTrap were
excluded from the analysis. Fig. 6 shows the total ion current
chromatograms (TICC) obtained from soy sauce A by vapor-
phase and immersion extraction and the peaks detected by
deconvolution. Here, 138 peaks were detected in vapor-phase
extraction, among which 123 components were identified by
the library search. On the other hand, 201 peaks were detected
in immersion extraction, and 147 hits were obtained. Based on
these results, in this experiment, the compounds in the soy
sauce samples could be extracted more comprehensively by
immersion extraction. Although low-boiling components such
as Isoamyl alcohol, Acetoin and Nonanal were common to both
vapor-phase extraction and immersion extraction, immersion
extraction allowed extraction of polar components and
relatively high boiling components, which rarely appear in the
vapor phase. The compounds that were qualitatively identified
in the library search included alcohols, furans, pyrazines, and
others, as well as characteristic aroma-active components of soy
sauces, namely, Homofuraneol (HEMF), p-Vinylguaiacol, and
Methionol. Table 2 shows the main compounds identified by
the library search following immersion extraction. Among these,
in some cases, the TIC included peaks and mass spectra that
could be extracted by deconvolution, even though they were
buried by other peaks, and the compounds could be identified
by a library search (Fig. 7). These results show that
comprehensive profiling of the main volatile components that
exist in a complex matrix is also possible even in scan
measurements by carrying out a deconvolution analysis.