User Benefits
Application
News
Table 1 Analytical Conditions and Target Compounds
Introduction
In the quality control of pharmaceuticals and foods with
functional claims, verifying the content of active ingredients and
detecting impurities are essential to ensure product efficacy and
safety. These tasks are typically performed by qualitative and
quantitative analysis using liquid chromatography (LC). However,
the large number of chromatograms generated must be reviewed
to identify peaks based on retention time, compare quantitative
results with specification limits, and confirm the presence or
absence of impurities. When these processes are performed
manually, they are susceptible to human error and the risk of
overlooking important findings. LabSolutions Detect
automatically analyzes the differences between a reference
chromatogram representing the impurity-free product and target
chromatograms obtained during routine quality control, thereby
preventing missed changes in active ingredient content and
overlooked impurity contamination. Application News No. 01-
01014 -EN introduced a case study in which LabSolutions Detect
was applied to the LC analysis of pramipexole hydrochloride, as
described in the USP, using the LC-2070C (UV model) to improve
the efficiency of quality control operations. In this article, a
method for further reducing the risk of missing impurities by
combining LabSolutions Detect with the LC-2070C 3D (PDA
model) and utilizing multi-wavelength information is presented.
LabSolutions Detect enhances the reliability and productivity of quality control operations by reducing the risk of human error
and oversight associated with visual inspection.
The use of a photodiode array (PDA) detector enables comprehensive detection across multiple wavelengths, further reducing
the risk of missing impurities.
Streamlining Quality Control Operations with
LabSolutions Detect—Reducing the Risk of
Missing Impurities with PDA Detector
Anomaly Peak Detection Support Software
Ultra High Performance Liquid Chromatograph
Shinichi Fujisaki
System : LC-2070C 3D
Control : Antipyrine and Probenecid
Sample : Control with added Salicylic acid
Mobile phase
Pump A : 0.1% formic acid in water
Pump B : Acetonitrile
Column : Shim-pack Scepter C18-120
*1
(150 mm × 4.6 mm I.D., 5 µm)
Injection Vol. : 5 µL
B Conc. : 20%(0 min)→95%(5-6 min)→20%(6-12 min)
Column Temp. : 40 ˚C
Flow rate : 1 mL/min
Detection : 261 nm
*1 P/N : 227-31020-05
Analytical Conditions and Target Compounds
Comprehensive Impurity Detection Using a
PDA Detector
By utilizing multi-wavelength data acquired with a photodiode
array (PDA) detector, LabSolutions Detect enables comprehensive
impurity detection. In addition, impurity peaks are automatically
detected according to user-defined acceptance criteria. In this
article, the upper and lower acceptance limits for the height of
unknown impurity peaks relative to the reference data were set to
10 mAU, and the target wavelength range was set to 190–400 nm
(Fig. 1). Under these conditions, impurity detection was
performed on the target data.
Fig. 1 Target Wavelength Range (Upper)
Acceptance Criteria for Unknown Impurity Detection (Lower)
The chromatograms obtained at a detection wavelength of 261
nm is shown in Fig. 2. For both the control (reference data) and
the sample (target data), only two compounds, Antipyrine and
Probenecid, were detected. Consequently, no unknown
impurities were detected in the target data when analysis was
performed at the single detection wavelength of 261 nm.
However, single-wavelength analysis carries the risk of
overlooking impurities that absorb outside the selected
wavelength. The contour display of PDA chromatograms
obtained from the multi-wavelength analysis of the sample
(target data) is shown in Fig. 3 (upper), where the vertical axis
represents the detection wavelength and the horizontal axis
represents the retention time. If a peak that deviates from the
specified criteria is detected, it is highlighted in red on the
contour display (red circle in the figure). The extracted
chromatogram obtained at a detection wavelength of 261 nm
and 303 nm is shown in Fig. 3 (lower). Although only Antipyrine
and Probenecid were detected at 261 nm, Salicylic acid was
detected as an unknown impurity at a detection wavelength of
303 nm (red circle in the figure). These results demonstrate that
the use of a PDA detector enables comprehensive impurity
detection across a wide wavelength range, thereby reducing the
risk of overlooking impurities.
Fig. 2 Chromatograms Obtained at a Detection Wavelength of 261 nm
Antipyrine
Probenecid
Control (Reference)
Sample (Target)
Detection: 261 nm
The samples and analytical conditions are summarized in Table 1.
An impurity-free control sample was used as the reference data,
while a control sample spiked with salicylic acid was used as the
target data.