Fig. 1 Chromatogram of mixed standard solution at 1 mg/L of chloride and
sulfate ions
Table 1 shows the analytical conditions, and Fig. 1 shows the
resulting chromatogram of mixed standard solution of chloride
and sulfate ions. The instrument used was a Nexera IC
suppressor-type ion chromatograph.
Linearity
Table 1 Analytical conditions
Table 2 Calibration levels and coefficients of determination (r
2
)
First-generation bioethanol produced from raw materials such as
corn and sugarcane has been utilized to achieve carbon neutrality.
However, research and development for second-generation
bioethanol produced from non-food, cellulose-based raw
materials is also underway to avoid competition with food crops.
In Japan, eight companies, including Toyota Motor Corporation,
have formed the “Research Association of Biomass Innovation for
Next Generation Automobile Fuels” (raBit) to advance research on
next-generation bioethanol technologies. Shimadzu participates
as a supporting member and promotes research and
development through analytical technology cooperation.
It is known that the presence of chloride and sulfate ions in fuel
ethanol can form engine residue, leading to reduced engine
performance. ASTM D4806-25
2)
specifies fuel ethanol quality as
criterion of chloride ion must not exceed 6.7 mg/kg and that of
sulfate ion must not exceed 4 mg/kg. ASTM D7319-22 includes
test methods for chloride and sulfate ions in fuel ethanol using
ion chromatography.
In this application, we present an example of the analysis of
chloride and sulfate ions in bioethanol using the Nexera IC in
accordance with ASTM D7319-22.
Introduction
Analysis of chloride and sulfate ions in fuel ethanol can be conducted in accordance with ASTM D7319-22
1)
.
Accurate quantitative determination, exhibiting excellent linearity and repeatability, can be achieved at concentration levels
significantly below the limits specified by the ASTM criteria.
Analysis of Chloride and Sulfate Ions in Bioethanol
Using the Nexera IC Ion Chromatograph
Tomoka Kaseda
Ion Chromatograph Nexera IC
System
:
Nexera IC
Column : Shim-pack IC-SA2
*1
(250 mm×4.0 mm I.D., 9 µm)
Guard Column : Shim-pack IC-SA28(G)
*2
(10 mm
×
4.6 mm I.D., 9 µm)
Mobile Phase : 1.8 mmol/L sodium carbonate
1.7 mmol/L sodium hydrogen carbonate
Flow Rate : 1.0 mL/min
Column Temp. : 30 ˚C
Injection Volume : 5 µL
Vial : Shimadzu Vial, LC, 4 mL Polypropylene
*3
Detection : Conductivity
Analysis of mixed standard solution
A calibration curve was created using seven calibration levels
for chloride ion and six for sulfate ion within the concentration
ranges specified in ASTM D7319-22. Table 2 shows the
respective calibration levels and coefficients of determination (r²)
for chloride and sulfate ions. It was confirmed that the r² values
of the calibration curves for both ions were 0.99 or more, as
specified in ASTM D7319-22.
Pretreatment
ASTM D7319-22 states that samples can be analyzed without
any pretreatment for chloride and sulfate ions in fuel ethanol.
However, a pretreatment of the addition of hydrogen peroxide
is required to analyze sulfur compounds (potential inorganic
sulfate) that are oxidized to sulfate ions in the presence of the
oxidant. The pretreatment method is shown in Fig. 2, and the
notation of ion species in accordance with ASTM D7319-22 is
shown in Table 3.
µS/cm
0.0 5.0 10.0 min
-0.10
0.00
0.10
0.20
0.30
15.0
Cl
SO
4
ions Calibration levels (mg/L) Coefficients of determination (r
2
)
Cl 0.3, 0.5, 1, 5, 10, 20, 50 ≥ 0.999
SO4 0.3, 0.5, 1, 5, 10, 20 ≥ 0.999
Sample 9.5 mL
Inject into the Ion chromatograph
30% Hydrogen peroxide 0.5 mL
Ion species Description Pretreatment
Total inorganic
chloride
Chloride ion without pretreatment
None
Total inorganic
sulfate
Sulfate ion without pretreatment
Potential
inorganic sulfate
Sulfate ion with oxidant
Addition of
hydrogen peroxide
Fig. 2 Pretreatment procedure (For potential inorganic sulfate measurement)
Table 3 Notation of ion species in accordance with ASTM D7319-22
Analysis of bioethanol samples and spike
and recovery test
Fig. 3 shows the analytical results for bioethanol sample and
standard-spiked bioethanol sample, to which a mixed standard
solution of chloride and sulfate ions was spiked to make
concentration of 1 mg/L for each in the sample solution. The
concentrations and recovery rates obtained for the spiked
sample are shown in Table 4. Calibration curves for quantitative
determination were prepared for both chloride and sulfate ions
using six levels less than 20 mg/L as shown in Table 2 to
consider the quantitative accuracy in the low-concentration
range.
*1 P/N: 228-38983-91 *2 P/N: 228-38983-92 *3 P/N: 228-31537-91