User Benefits Application News Whole blood for trace element analysis can be easily contaminated during sampling and storage. Thus, special attention needs to be paid to all collection devices, including needles, stoppers, lubricants and separator systems, as even µg/L-levels can bias results 1) . Usually, samples are immediately transferred into pre-cleaned, trace-metal–certified polypropylene or PMP tubes to minimize contact with glass and stainless-steel surfaces that can leach chromium, nickel, antimony or other elements. To prevent clotting, laboratories employ anticoagulants added in precisely defined blood-to-additive ratios 1) . It should be well noted, which kind of anticoagulant has been used. Depending on anticoagulant, an extraordinary portion of e.g. potassium, sodium, lithium or other components is added to the blood, which might affect the analysis. Instead of laborious and time-consuming sample preparation, e.g., by microwave-assisted acid digestion, this approach is simple dilution. Because protein-rich matrices can be unstable or cause clogging when acidified, an alkaline dilution strategy (Table 1) provides the robustness and element stability needed for accurate whole-blood analysis while offering improved mercury washout capabilities. The same alkaline dilution strategy was successfully applied in recent ICP-MS Application News to urine [1.] and serum [2.]. Sample Preparation and Calibration For matrix matched calibration, Whole Blood Calibrator (ClinCal® 9943, Recipe) was injected in different dilution levels (cf. Table 2). For quality control like proving stability and sensitivity characteristics of the method, Whole Blood Controls Level I, II, III (ClinCheck® 8840-8843, Recipe) were measured frequently throughout the batch. Controls as well as samples are simply 20-fold diluted by alkaline diluent and well homogenized (vortex mixing) prior to analysis. Inductively coupled plasma mass spectrometry (ICP-MS) like Shimadzu ICPMS-2050 (Fig. 1) offers ultra-low detection limits and fast multi-element capability required for reliable high throughput measurement of nutritional, essential, and toxic elements in various complex biological samples. While urine has been widely used to monitor elemental excretion, blood serum represents the protein-free fraction of circulating blood. Whole blood, however, contains both cellular and extracellular components, enabling a more integrated assessment of an individual’s nutritional status and exposure to potentially toxic metals than either urine or serum alone. Accordingly, whole-blood ICP-MS analysis is well suited for research applications such as nutritional assessment, toxicology, and biomonitoring studies. Introduction Offering gas switching times <5 sec and ProActive Rinsing function, short sample cycle times can be achieved Using whole blood calibrators and control samples in research labs simplifies working routines Alkaline media drastically improves washout capabilities for mercury and ensures proper sample stabilization compared to acidic media Determination of More Than 25 Nutritional, Essential and Toxic Elements in Whole Blood by ICP-MS Using Alkaline Dilution Jan Knoop Inductively Coupled Plasma Mass Spectrometry ICPMS-2050 LF Shimadzu Europa GmbH Fig. 1 ICPMS-2050 LF with AS-20 Autosampler Table 1 Components of Alkaline Diluent prepared in Water *1 Reagent Description Purpose Ammonia 25% (v/v), Suprapur®, Merck, Germany adjusts alkaline pH Tergitol 15-S-9 extra pure, Carl Roth, Germany Homogenizes sample, supports lyse of cells EDTA (Na-free) >99.999% Sigma-Aldrich, USA stabilizes ions in solution Isopropanol ≥99.999%, Sigma-Aldrich, USA compensates carbon enhancement effect *2 *1 All solutions are prepared using ASTM Type 1 ultrapure water (18.2 MΩ resistivity, Arium® Pro VF, Sartorius AG). *2 Carbon present in sample will otherwise increase response of less ionized elements (e.g. As, Se), resulting in false positives. Element, unit Cal 1 Blank Diluent Cal 2 12.5% Cal 3 25% Cal 4 50% Cal 5 100%* Ag µg/L 0 1.04 2.60 5.20 10.4 Al µg/L 0 2.96 7.40 14.8 29.6 As µg/L 0 1.46 3.65 7.30 14.6 Ba µg/L 0 3.93 8.83 19.7 39.3 Ca mg/L 0 4.80 12.0 24.0 48.0 Cd µg/L 0 1.38 3.45 6.9 13.8 Co µg/L 0 1.49 3.725 7.45 14.9 Cu mg/L 0 0.171 0.42750 0.855 1.71 Fe mg/L 0 38.2 95.5 191 382 Hg µg/L 0 3.59 8.98 18.0 35.9 I µg/L 0 9.49 23.7 47.5 94.9 K mg/L 0 236 590 1180 2360 Mg mg/L 0 4.36 10.9 21.8 43.6 Mn µg/L 0 2.59 6.48 13.0 25.9 Mo µg/L 0 1.04 2.60 5.20 10.4 Na mg/L 0 206 515 1030 2060 P mg/L 0 37.0 92.5 185 370 Pb µg/L 0 34.7 86.8 174 347 Pd µg/L 0 0.659 1.65 3.30 6.59 Pt µg/L 0 0.673 1.68 3.37 6.73 Sb µg/L 0 1.68 4.20 8.40 16.8 Se µg/L 0 30.8 77.0 154 308 Sn µg/L 0 1.07 2.68 5.35 10.7 Sr µg/L 0 6.78 17.0 33.9 67.8 Tl µg/L 0 1.06 2.65 5.30 10.6 Zn mg/L 0 0.828 2.07 4.14 8.28 Table 2 Recipe® Calibrator in alkaline diluent, µg/L * 100% corresponds to 0.5 mL in 10 mL (considering a 20-fold dilution)
Application News Configuration and Measurement Conditions Table 3 ICPMS-2050 LF Configuration and Basic Method Settings Autosampler : AS-20 (Covered Round Table / Dual Rinse / Spacer for low dead volume) Extended Cover : ENC-280 (in case dust is an issue) Sample Probe : Inert Guardian Probe with inline filter Spray Chamber : Cyclonic Twister (reduces carry-over) Chamber Temperature : 5 °C Nebulizer : Coaxial DC 04 (~0.4 mL/min) Peristaltic Pump : 4 Channels, 12-roller @ 20 rpm Internal Standard Tubing : PVC, flared-end, orange-blue Sample Tubing : PVC, flared-end, black-black Drain : Automatic overflow type Torch : Shimadzu Minitorch Cone Material : Ni Sampling Depth : 9.0 mm RF-Power : 1.20 kW Plasma Gas : Ar 9.0 L/min Auxiliary Gas : Ar 1.10 L/min Carrier Gas : Ar 0.47 L/min Dilution Gas : Ar 0.38 L/min Cell Gas : He 6.0 mL/min H 2 7.0 mL/min Cell Voltage : -21.0 V Energy Filter : 7.0 V System configuration of ICPMS-2050 LF was selected according to Table 3. All components have been selected with care to fulfill body fluid analysis requirements. More than 25 elements have been selected for quantification (Table 4) whereat some of the selected isotopes are partwise affected by polyatomic interferences. To eliminate interferences, ICPMS-2050 LF highly efficient He-collision mode, H 2 -reaction mode as well as high- resolution mode was applied. Element Isotope Integration time [s] Cell Gas Internal Standard Ag 107 0.4 He 103 Rh Al 27 5.0 H2 45 Sc As 75 1.0 He 89 Y Ba 137 0.4 He 103 Rh Ca 44 0.1 He 45 Sc Cd 111 2.0 He 103 Rh Co 59 0.5 He 45 Sc Cu 63 0.1 He 89 Y Fe 57 0.1 He 45 Sc Hg 202 0.3 He 193 Ir I 127 0.2 He 103 Rh K 39 0.1 He 45 Sc Mg 25 0.1 He 45 Sc Mn 55H* 1.0 He 45 Sc Mo 95 0.5 He 89 Y Na 23 0.1 He 45 Sc P 31 0.1 NoGas 45 Sc Pb 206+207+208 0.3 He 193 Ir Pd 105 0.5 He 89 Y Pt 195 0.3 He 193 Ir Sb 121 0.5 He 103 Rh Se 78 0.3 H2 89 Y Sn 118 0.5 He 103 Rh Sr 88 0.3 He 89 Y Tl 205 0.2 He 193 Ir Zn 66 0.1 He 89 Y *measured in high resolution mode Table 4 Elements and Isotopes Results and Discussion Method Stability The method was stable over the full 100 samples runtime indicated by internal standard recovery, standardized to the lowest calibration curve blank level (Fig. 3). At the same time, whole blood control LI-LIII were frequently measured and have been satisfactorily traced back to given values within specified reference ranges. The reference values may be entered to the LabSolutions ICPMS method directly, allowing for quick assessment of the quality of the results obtained from each run, as outliers will be highlighted automatically. The software also allows to easily monitor internal standard stability. Fig. 3 Internal Standard Stability Plot 0% 20% 40% 60% 80% 100% 120% 140% 0:00 0:26 0:50 1:15 1:40 2:05 2:29 2:54 3:19 3:44 4:09 4:34 4:58 5:23 Recovery Rate (Cal-Blank = 100%) Time [h:min] Ir 193 (He) Rh 103 (He) Sc 45 (NoGas) Sc 45 (He) Sc 45 (H2) Y 89 (He) Y 89 (H2) 125% 60% Fig. 2 Calibration Curves Alkaline diluent was measured as blank sample. All calibration curves (Fig. 2) are weighted inversely (1/I) to improve the precision in the low-end trace region.
Application News www.shimadzu.com/an/ Shimadzu Corporation Copyright © 2026 Shimadzu Corporation and/or its affiliates. All rights reserved. For Research Use Only. Not for use in diagnostic procedures. This publication may contain references to products that are not available in your country. Please contact us to check the availability of these products in your country. The content of this publication shall not be reproduced, altered or sold for any commercial purpose without the written approval of Shimadzu. See http://www.shimadzu.com/about/trademarks/index.html for details. Third party trademarks and trade names may be used in this publication to refer to either the entities or their products/services, whether or not they are used with trademark symbol “TM” or “”. Shimadzu disclaims any proprietary interest in trademarks and trade names other than its own. The information contained herein is provided to you "as is" without warranty of any kind including without limitation warranties as to its accuracy or completeness. Shimadzu does not assume any responsibility or liability for any damage, whether direct or indirect, relating to the use of this publication. This publication is based upon the information available to Shimadzu on or before the date of publication, and subject to change without notice. SHIMADZU Europa GmbH, www.shimadzu.eu First Edition: Sep. 2026 05-SCA-116-014-EN Conclusion In general, the method is developed to be stable while allowing to analyze the usual reference ranges (Table 5). The sensitivity is determined by standard deviation of the calibration curve blank sample. Because samples and reference samples are treated in the same way, the dilution factor can be disregarded, allowing to use the given limit of quantification (LOQ) as approximation for the method quantification limit (MQL). The ICPMS-2050 LF was specifically configured for robust small volume analysis while reducing carry over, sample cycle times and environmental contamination impacts while being capable of successfully addressing several kind of possibly interference providing most accurate and reliable results. The collision- and reaction cell performs efficiently with flowrates of helium and hydrogen in low levels of 6-7mL/min. At the same time, the method uses the well established Minitorch plasma system, allowing lowest argon consumption while reducing power consumption, which makes it exceptionally cost-effective without compromising performance characteristics. LabSolutions ICPMS software allows to easily control the device, consumable lifetime, QC-recoveries and internal standard stability. Furthermore, features like ProActive Rinsing and Extended Rinsing speed up the method while reducing carry- over risks. <References> 1) Blood collection device components: issues, innovations, and recommendations for clinical laboratories and manufacturers – a narrative review, Journal of Laboratory and Precision Medicine, Vol 10 (2025) https://jlpm.amegroups.org/article/view/10316/html 2) Inductively Coupled Plasma Mass Spectrometry: Introduction to Analytical Aspects, Clin Biochem Rev 40 (3) 2019, p. 126. 3) Medical laboratory Bremen, https://www.mlhb.de/analysen (accessed 24.04.2026) <Related Applications> 1. Determination of 23 Nutritional, Essential and Toxic Elements in Urine by ICP-MS Using Alkaline Dilution, Application News 05-SCA-116-013-EN 2. Determination of 20 Nutritional, Essential and Toxic Elements in Blood Serum by ICP-MS using Alkaline Dilution, Application News 05-SCA-116-010-EN 3. ICP-MS Analysis of 25 Essential, Trace and Nutritional Elements in Blood Serum Using Recipe ClinMass® Complete Kit, Application News 05-SCA-116-012-EN Method Sensitivity Element, unit LOD (3s) ≈ MDL LOQ (10s) ≈ MQL Common Reference Range 3) Recipe Control Range (LI – LIII) ICPMS- 2050 LF suitable? Ag µg/L 0.02 0.05 < 0.6 1.55-7.55 ● Al µg/L 0.95 3.18 < 20.5 12.0-26.9 ● As µg/L 0.05 0.15 < 3.4 3.1-10.8 ● Ba µg/L 0.48 1.61 < 2.9 8.71-25.3 ● Ca mg/L 0.46 1.52 44.9-63.4 45.3-46.3 ● Cd µg/L 0.05 0.17 < 1.0 1.59-6.98 ● Co µg/L 0.04 0.13 < 0.9 1.45-12.7 ● Cu mg/L 0.0002 0.0006 0.65-1.31 0.758-1.51 ● Fe mg/L 0.18 0.61 378-600 376-385 ● Hg µg/L 0.09 0.31 < 2.0 2.31-18.5 ● I µg/L 0.53 1.75 - 25.7-59.8 K mg/L 0.30 0.99 1485-1955 1260-2050 ● Mg mg/L 0.02 0.08 15-26 (serum) 26.8-38.2 ● Mn µg/L 0.29 0.96 5.0-13.5 8.68-6.51 ● Mo µg/L 0.09 0.29 0.2-1.3 2.00-8.17 ● Na mg/L 0.46 1.54 1563-1954 1960-1980 ● P mg/L 0.68 2.27 314-443 345-353 ● Pb µg/L 0.08 0.27 < 30 35.0-264 ● Pd µg/L 0.03 0.10 < 0.2 1.2-4.98 ● Pt µg/L 0.02 0.05 < 0.2 (serum) 1.67-5.06 ● Sb µg/L 0.02 0.08 < 0.2 2.10-8.39 ● Se µg/L 0.06 0.20 73-165 88.2-204 ● Sn µg/L 0.18 0.59 < 2.0 2.59-8.05 ● Sr µg/L 0.19 0.63 9.0-41.0 15.3-48.6 ● Tl µg/L 0.07 0.23 < 0.6 0.868-8.68 ● Zn mg/L 0.001 0.005 4.0-7.8 4.83-7.77 ● ● means the method sensitivity is in line with common reference ranges. Table 5 ICPMS-2050 LF Method Sensitivity [µg/L] To further increase sensitivity and accuracy for individual elements, a simple approach can be to increase its integration times, which is recommended for Molybdenum, Palladium and Tin. A usual troublesome element is Selenium: “From a clinical biochemistry point of view, perhaps the most significant example is the interference of gadolinium with selenium. […] Gadolinium isn’t normally present in biological samples; however, it is present in some intravenous contrast agents used for magnetic resonance imaging (MRI) […]. The presence of gadolinium in a clinical sample can cause an artefactually elevated selenium result” 2) , as double charged 156 Gd interferes 78 Se. As it was shown within the related application note [2.], Gd-interferences can be sufficiently removed by applying H2- reaction mode (Table 6). Eliminate Possible Interferences Solution A Solution B Solution C Solution D Gd-level 0 ppm 10 ppm 20 ppm 50 ppm Se-level 20 ppb 20 ppb 20 ppb 20 ppb Se-Result H 2 -mode 9.9 ppb 19.7 ppb 19.7 ppb 19.8 ppb Table 6 Eliminated effect of 156 Gd to 78 Se in H 2 -mode LabSolutions is a trademark of Shimadzu Corporation or its affiliated companies in Japan and/or other countries. Another issue to be addressed in whole blood is the accurate recovery of trace Manganese levels ( 55 Mn, ~10-20ppb), being partially influenced by the neighboring dominant iron signal ( 56 Fe, ~400ppm). This is not an issue in case the iron-content is stable and matrix matched calibration is applied. But in case the Fe-level varies in sample or was not considered in creating the curve, special attention needs to be paid to avoid false positive or false negative Mn-results, resp. To be independent from variations in iron levels, 55 Mn was selected to be measured in high resolution mode while applying Inter Element Correction (IEC), which ensures any remaining interfering levels are determined and corrected with each sample individually, allowing to achieve most accurate Mn-results. In Table 7 it is shown that even 500ppm iron does no further result in false-positives. Calibration Solution A Solution B Solution C Fe-level 0 ppm 500 ppm 500 ppm 500 ppm Mn-level 0-50 ppb 0 ppb 10 ppb 50 ppb Mn-result - <0.3 ppb 9.4 ppb 48.4 ppb Table 7 Eliminated effect of 56 Fe to 55 Mn
Please fill out the survey Related Products Some products may be updated to newer models. ICPMS-2040 Series / ICPMS-2050 Series Inductively Coupled Plasma Mass Spectrometry Related Solutions Clinical Research and Forensics Clinical Research Life Science Biomarker Price Inquiry Product Inquiry Technical Service / Support Inquiry Other Inquiry Powered by TCPDF (www.tcpdf.org)