NOA-7100 - Specs

Transportable NOx-O₂ Gas Analyzer

Standard type

Two standard types are available.
 

  Type 1 Type 2
Measurement components NOx, O2 NOx, NO, NO2 (*1)
Measurement range NOx: 0-25/50/100/250/500/1000/2500/4000 ppm
O2: 0-5/10/25 vol%
NOx, NO, NO2:
0-100/250/500/1000/2500/4000 ppm (*2)
Measurement principle NOx: Atmospheric pressure chemiluminescence method; O2: Limiting current type zirconia method
Repeatability Within ±0.5% of full scale
Zero drift Within ±1% of full scale per day (25 ppm range: within ±1.5% of full scale per day)
Span drift Within ±1.5% of full scale per day
Linearity Within ±2% of full scale
Response time (*3) NOx: Approx. 20/30/60 sec (Switching is configurable)
O2: Approx. 30/60 sec (Switching is configurable)
NOx/NO: Within 90 sec
Gas sample amount Approx. 2 L/min 100 mL/min
Display Liquid crystal display, able to display 3 items at a time
Transmission output 0-1VDC, 3 channels, insulated output (individual channels not insulated)
Wireless data transmission Equipped
Data output to external media (*4) Measurement data stored in the internal memory of the analyzer can be written to USB flash drives
in CSV file format; can also be output to a computer via Wi-Fi or LAN cable connection
Permissible ambient temperature (*5) 5-40°C; must not be exposed to direct sunlight or radiant heat
Permissible ambient humidity (*5) Relative humidity 90%RH or less
Power requirements Choose from 110, 115, 120, 127, 220, 230, 240 VAC 50-60 Hz, 100 V series: 410 VA, 200 V series: 700 VA (steady 200 VA)
Dimensions W260 × H452 × D420 mm (does not include protrusions)
Weight (analyzer) Approx. 16 kg
External drain separator Fitted Not fitted
Sample gas conditions (*6) [ At NOA-7100 inlet port ] Pressure: -980 to + 980 Pa; Temperature: Ambient temperature; Dust: 0.1 g/Nm3 or less
External signal input 0-1VDC, 1 channel, Signal input from an external analyzer, display (Data can be saved)
Ambient temperature

The following specifications are the same as for the standard model.
Display, transmission output, permissible ambient temperature, permissible ambient humidity, power requirements, dimensions, weight (body), sample gas conditions, external signal input
*1: NO2 is measured by calculation (NOx-NO).
*2: NOx, NO, and NO2 are in the same range. The maximum NO2 concentration is 300 ppm.
*3: Time that indicates 90% response from gas flowing from sample gas inlet (flowing gas switching time + electrical system
response time).
      With Type 2, this will be the response when NOx-NO switching is fixed.
*4: Check the Shimadzu website for smart devices and USB flash drives for which operation has been confirmed.
*5: Permissible ambient temperature and humidity are indicated by the shaded part of the figure.
*6: Sample gas conditions and coexisting gases are summarized below. Please contact your Shimadzu representative if the conditions are not satisfied.

{"title":"Specification Sheet Downloads","description":"Download the latest specification sheets.","source":"product","key":3498,"max":30,"filter_types":["spec_sheet"],"link_title":"View other Downloads","link_url":""} {"title":"Downloads","description":"Download the latest brochure.","source":"product","key":3498,"max":30,"filter_types":["brochures"],"link_title":"View other Downloads","link_url":"","pdf_links":[]}