OPTIC-4 - Specs

Multimode Inlet for Gas Chromatograph-Mass Spectrometer

Inlet

Maximum Usage Temperature

600 °C (GC oven temperature of 35°C)

Coolant Temperature

35 °C (air)

Heating Speed

1 to 60 °C/sec

Thermal Program

9 step

Read more
Show Less

Electronic Flowrate Control

Split line equipped with a solvent sensor

Pressure Range

7 to 700 kPa*

Total Flowrate Range

5 to 500 mL/min**

Pressure Sensor

Precision ±1 % (full scale)
Repeatability ±0.2 % (full scale)

Flowrate Sensor

Precision ±1 % (full scale)
Repeatability ±0.2 % (full scale)

* Values of AFC-2010/AFC-2030 : 0 to 970 kPa
** Values of AFC-2010 : 0 to 1200 mL/min,Values of AFC-2030 : 0 to 1300 mL/min

Read more
Show Less

Cryotrap (Option)

Operating temperature range

-150 to +350 °C

Heating Speed

1 to 60 °C/sec

Coolant

Liquid nitrogen(pressurized container 150 to 200 kPa)

Installation Environment

Dimensions

W 140 × D 340 × H 340 mm

Weight

8.4 kg (control unit)

Ambient temperature

18 to 40 °C

Ambient Humidity

40 to 70 %

Main Power

100 to 240 VAC, 50/60 Hz

Power Consumption

150 W (normal), 450 W (maximum)

Read more
Show Less
{"title":"Specification Sheet Downloads","description":"Download the latest specification sheets.","source":"product","key":4491,"max":"30","filter_types":["spec_sheet"],"link_title":"View other Downloads","link_url":""} {"title":"Downloads","description":"Download the latest brochure.","source":"product","key":4491,"max":"30","filter_types":["brochures"],"link_title":"View other Downloads","link_url":"","pdf_links":[]}