PPSQ-51A/53A - Configuration
Protein Sequencer
Other Items Provided by User
- Isocratic System
Installation Space Tabletop: W1600 x D600 x H800 mm min. Weight: Approx. 100 kg Power Supply 100 V AC, 15 A 50/60 Hz Nitrogen Gas Purity: 99.9999% min. A cylinder pressure regulator and gas piping (10 m)
are included as standard.Exhaust Equipment To vent vapors from waste liquids, an exhaust tube must be connected to exhaust equipment, or vented outdoors. An exhaust tube (20 m) is included as standard. - Gradient System
Installation Space Lab Bench: W1800 x D600 x H800 mm min. Weight: Approx. 120 kg Power Supply 100 V AC, 15 A 50/60 Hz Nitrogen Gas Purity: 99.9999% min. A cylinder pressure regulator and gas piping (10 m)
are included as standard.Exhaust Equipment To vent vapors from waste liquids, an exhaust tube must be connected to exhaust equipment or vented outdoors. An exhaust tube (20 m) is included as standard.
Note: : The installation space and the power requirements may vary depending on the PC and monitor used.
Installation Example
Isocratic System