i-Series - Specs

High Performance Liquid Chromatograph

LC-2050C (UV model)

Pump

Degassing unit

Five Lines: Mobile phase 4 + Rinse solution 1 [Volume 400 μL]

Pumping method

Parallel-type double plunger

Pulsation

≤ 0.1 MPa [1.0 mL/min,10 MPa, Water]

Flow rate setting range

0.0001 to 10 mL/min

Flow rate accuracy

≤ ±1% or ≤ ±2 μL/min, below whichever is greater
[0.01 to 2 mL/min, Specified condition]
≤ ±2%
[2 to 5 mL/min, Specified condition]

Flow rate precision

< 0.06 %RSD or < 0.02 minSD, below whichever is greater

Configuration

Four-solvent low-pressure gradient

Gradient / range of set concentrations

0 to 100%, in 0.1% steps

Gradient / concentration accuracy

±0.5% [0.1 to 2 mL/min, 1 to 20 MPa, Specified condition]

Gradient / concentration precision

±0.1% [1 mL/min,10 MPa, Specified condition]

Maximum pressure

50 MPa [0.0001 to 3 mL/min]
44 MPa [3.0001 to 5 mL/min]
22 MPa [5.0001 to 10 mL/min]

System Delay Volume

650 μL
[Option: 460 μL, 1100 μL]

Read more
Show Less

Autosampler

Sample injection method

Total-volume sample injection

Injection volume accuracy

±1% [50 μL, N = 10]

Injection volume setting range

0.1 to 100 μL
[Option: 0.1 to 50 μL, 1 to 500 μL, 1 to 2,000 μL]

Injection volume reproducibility

RSD < 0.20% [5.0−2000 μL]
RSD < 0.25% [2.0−4.9 μL]
RSD < 0.5% [1.0−1.9 μL]
RSD < 1.0% [0.5−0.9 μL]

Cross-contamination

0.0025% [Caffeine, Specified condition]

Injection cycle time

Min. 14 sec [Specified condition]

Samples for processing

336 [1 mL], 216 [1.5 mL], 112 [4 mL], 4 [MTP/DWP]

Sample cooler

4 to 45°C
[Room temperature needs to be 30°C or lower and the humidity 70% or less to go down to 4°C. Room temperature needs to be 15°C or higher to go up to 45°C.]

Injection linearity

> 0.9999
[1 to 100 μL, Specified condition]

Read more
Show Less

Column Oven

Heating and cooling method

Forced air circulation method

Containable column size

Six columns 10 cm long
Three columns 30 cm long
One 30 cm long column if a CTO unit (optional) for i-Series low-temperature
analysis is installed. (Two columns up to 15 cm long or one column 30 cm long
if an FCV tubing kit for i-Series low-temperature analysis is also installed.)

Temperature control range

Room temperature −12 °C (−15 °C if an optional unit is installed) to 90 °C

Temperature control precision

±0.1°C

Temperature accuracy

±0.8°C (Specified condition)

Flow rate switching valve

Max. 1 pc

Read more
Show Less

UV Detector

Wavelength range

190 to 700 nm

Spectral bandwidth

8 nm

Wavelength accuracy

≤ ±1 nm

Wavelength Reproducibility

≤ ±0.1 nm

Noise

≤ ±2.5 × 10−6 AU, (250 nm, Specified condition)

Drift

≤ 100 × 10−6 AU/h [250 nm, Specified condition]

Simultaneous monitoring of 2 wavelengths

Enable [Any 2 wavelengths of 190 to 370 nm or 371 to 700 nm]

Linearity

Up to 2.5 AU (5%)

Sampling rate

Up to 100 Hz

Light source

Deuterium (D2) lamp

Flow Cell

12 μL (10 mm, TC), 12 MPa

Option cell

High-Speed: 8 μL [10 mm, TC]
Semi-micro: 2.5 μL [5mm, TC]

Read more
Show Less

Miscellaneous

Dimensions

W410×H605×D500 mm [Not including reservoir tray height]

Weight

63 kg

Available pH range

1 to 13

Materials for parts in contact with liquids

Stainless steel (SUS316L, SUS316), FEP, PEEK, PTFE, perfluoroelastomer, ruby, sapphire, Hastelloy® C, GFP, ceramic, PFA, quartz, PPS

Read more
Show Less
{"title":"Specification Sheet Downloads","description":"Download the latest specification sheets.","source":"product","key":4554,"max":"30","filter_types":["spec_sheet"],"link_title":"View other Downloads","link_url":""} {"title":"Downloads","description":"Download the latest brochure.","source":"product","key":4554,"max":"30","filter_types":["brochures"],"link_title":"View other Downloads","link_url":"","pdf_links":[]}