TOC-1000e S - Specs
On-Line TOC Analyzer for Ultrapure Water
Product specification
| Measurement items | TOC |
|---|---|
| TOC measurement principle | UV oxidation-conductivity measurement method |
| Measurement range*1 | 0 to 500 μg/L |
| Detection limit*1 | 0.02 μg/L |
| Repeatability*1 | ±0.05 μg/L (when concentration is no greater than 1 μg/L) |
| Accuracy*1 | ±5% (500 μg/L sucrose) |
| Linearity*1 | R2 = 0.98 or greater (when measuring 250 mg/L in the 0-500 mg/L range) |
| Startup time | 5 minutes |
| Measurement cycle | 5 minutes, 10 minutes, 15 minutes, 30 minutes |
| External dimensions | 270 (W) × 180 (H) × 140 (D) mm |
| Required power supply | 100 to 240 V AC, 50/60 Hz 100 VA |
| Weight | 約3.0 kg |
| Protection from dust and water spray | IP33 |
| Wiring length | 10 m or less (when using LAN, analog, and contact) |
- *1 The specifications apply when the oxidation time is set to 70 seconds.
Sample conditions and Installation environment*2
| Sample temperature*3 | 25±5 °C (Guaranteed operation: 10 to 50 °C) |
|---|---|
| Sample conductivity | 0.1 μS/cm or less (Guaranteed operation: 1 μS/cm or less) (both at 25 °C sample temperature) |
| Sample flow rate | 30 to 200 mL/min (Recommended: 100 mL/min, guaranteed operation: 30 to 500 mL/min) |
| Ambient temperature*3/humidity | 25±5 °C (Guaranteed operation: 10 to 40 °C) 0 to 80 % (No condensation) |
- *2 The samples and environmental conditions must be within the specifications and the changes should be small.
*3 The difference between the ambient temperature and the sample temperature should be small.