English
Español
English
0

    Nexera Prep - Specs

    NEW

    Preparative Purification Liquid Chromatograph

    Excellent System Expandability

    Solvent Delivery Unit Accommodates a Wide Range of Recovery Volumes

    Guidelines for Preparative Scale and Maximum Load

    Guidelines for Preparative Scale and Maximum Load

    The figure at right shows the guidelines for total component capacity with a 250 mm long column when the target component is highly soluble in the mobile phase, separates from impurities, and ions are suppressed. For isocratic elution, in principle, the total component capacity is proportional to the column volume.

    LC-40D
    Supports a Range of Applications from High-Precision Analytical to Semi-Preparative

    LC-40D

    LC-40D

    • This solvent delivery unit can handle flow rates ranging from those used in analytical scale to those used in semi-preparative (up to 10 mL/min).
    • High-precision analysis is possible even in the semi-micro flow-rate range.

    LC-20AR
    Supports Semi-Preparative and Recycle Preparative

    LC-AR

    LC-20AR

    • This solvent delivery unit can handle flow rates used in semi-preparative scale (up to 20 mL/min).
    • Using a recycle kit enables semi-preparative recycling.

    LC-20AP
    Supports Large-Scale Preparative Fractionation

    LC-20AP

    LC-20AP

    • High flow rates (up to 150 mL/min) enable highly efficient, large-scale preparative fractionation.
    • Large-scale prep solvent delivery fully supports the preparative fractionation workflow, including reinjection, to assess purity.
    • Combine with an FCV-200AL low-pressure gradient unit to perform gradient analysis using up to four mobile phases.
    Specifications
      LC-40D LC-20AR LC-20AP
    Solvent Delivery Method Parallel-type double plunger
    Plunger Capacity 10 μL 47 μL 250 μL
    Maximum Discharge Pressure 44 MPa 49 MPa 42 MPa
    Flow Rate Setting Range 0.0001 to 5.0000 mL/min (1.0 to 44 MPa)
    5.0001 to 10.0000 mL/min (1.0 to 22 MPa)
    0.01 to 20.00 mL/min 0.01 to 150.00 mL/min
    Flow Rate Accuracy No more than ±1% or ±2 μL/min, whichever
    is greater (under specified conditions)
    No more than ±1% or ±10 μL/min,
    whichever is greater (0.1 to 5.0 mL/min
    No more than ±1% (1 mL/min, 10 MPa)
    Flow Rate Precision No more than 0.06% RSD or 0.02 min SD, whichever is greater No more than 0.08% RSD or 0.02 min SD, whichever is greater No more than 0.1% RSD or 0.02 min SD, whichever is greater
    Constant Pressure Solvent Delivery Supported
    Plunger Rinsing Mechanism Optional available Syringe or rinsing pump (228-39625-41) Syringe or rinsing pump (228-39625-41)
    Operating Temperature Range 4 to 35°C
    Size and Weight W260 × H140 × D500 mm, 10 kg W260 × H140 × D500 mm, 16 kg W260 × H210 × D500 mm, 19 kg

    LC-40D

    LC-40D

    LC-40D

    • This solvent delivery unit can handle flow rates ranging from those used in analytical scale to those used in semi-preparative (up to 10 mL/min).
    • High-precision analysis is possible even in the semi-micro flow-rate range.

    Guidelines for Preparative Scale and Maximum Load

    Solvent delivery method

    Parallel-type double plunger

    Plunger capacity

    10 μL

    Maximum discharge pressure

    44 MPa

    Flow rate setting range

    0.0001 to 5.0000 mL/min (1.0 to 44 MPa)
    5.0001 to 10.0000 mL/min (1.0 to 22 MPa)

    Flow rate accuracy

    No more than ±1% or ±2 μL/min, whichever
    is greater (under specified conditions)

    Flow rate precision

    No more than 0.06% RSD or 0.02 min SD, whichever is greater

    Constant pressure solvent delivery

    Supported

    Plunger rinsing mechanism

    Optional available

    Operating temperature range

    4 to 35 °C

    Size and weight

    W260 × D500 × H140 mm, 10 kg

    Read more
    Show Less

    LC-20AR

    7 - Nexera Prep

    This solvent delivery unit can handle flow rates used in semi-preparative scale (up to 20 mL/min).
    Using a recycle kit enables semi-preparative recycling.

    Supports Semi-Preparative and Recycle Preparative

    Solvent delivery method

    Parallel-type double plunger

    Plunger capacity

    47 μL

    Maximum discharge pressure

    49 Mpa

    Flow rate setting range

    0.001 to 20.000 mL/min

    Flow rate accuracy

    No more than ±1% or ±10 μL/min,whichever is greater (0.1 to 5.0 mL/min)

    Flow rate precision

    No more than 0.08% RSD or 0.02 min SD, whichever is greater

    Constant pressure solvent delivery

    Supported

    Plunger rinsing mechanism

    Syringe or rinsing pump

    Operating temperature range

    4 to 35 °C

    Size and weight

    W 260 x H140 x D 500 mm,16 kg

    Read more
    Show Less

    LC-20AP

    7 - Nexera Prep

    High flow rates (up to 150 mL/min) enable highly efficient, large-scale preparative fractionation.
    Large-scale prep solvent delivery fully supports the preparative fractionation workflow including reinjection to assess purity.
    Combine with an FCV-200AL low-pressure gradient unit to perform gradient analysis using up to four mobile phases.

    Supports Large-Scale Preparative Fractionation

    Solvent delivery method

    Parallel-type double plunger

    Plunger capacity

    250 μL

    Maximum discharge pressure

    42 Mpa

    Flow rate setting range

    0.01 to 150.00 mL/min

    Flow rate accuracy

    No more than ±1% (1 mL/min, 10 MPa)

    Flow rate precision

    No more than 0.1% RSD or 0.02 min SD,whichever is greater

    Constant pressure solvent delivery

    Supported

    Plunger rinsing mechanism

    Syringe or rinsing pump

    Operating temperature range

    4 to 35 °C

    Size and weight

    W 260 x H210 x D 500 mm,19 kg

    Read more
    Show Less

    Shim-pack Scepter Columns

    - Excellent stability & performance using a wide range of LC conditions -

    Shim-pack Scepter LC columns, which are the next generation organic silica hybrid based columns, are designed for stability and performance in a wide range of mobile phase conditions. With different chemistry characteristics, Shim-pack Scepter columns are effective for method development/scouting under conditions that may compromise traditional silica based columns.
    With different particle sizes (1.9 μm, 3 μm, 5 μm) and different column dimensions, Shim-pack Scepter LC columns are fully scalable between UHPLC, HPLC and preparative LC, making method transfer seamless between different laboratory instrumentation.

    Shim-pack Scepter Reversed Phase
    C18 HD-C18 C8 Phenyl PFPP
    Functional Group Trifunctional C18 Trifunctional C18 Trifunctional C8 Trifunctional
    Phenylbutyl
    Trifunctional
    Pentafluorophenylpropyl
    Generic Purpose Type High Density Type
    Particle Organic Silica Hybrid
    Particle Size 1.9 μm,3 μm,5 μm
    Pore Size 12 nm 8 nm 12 nm
    End Capping Proprietary None
    pH Range 1 - 12 1 - 10 1 - 8
    100% Aqueous Condition Yes No No Yes Yes
    USP Classification L1 L1 L7 L11 L43