Solvent Delivery Pump

 

LC-40B XR

 

LC-40B X3

  LC-40D
 
LC-40D XR
LC-40B XR
LC-40D XS LC-40D X3
LC-40B X3
Pumping method Parallel-type double plunger (approx. 10 μL/1 stroke)
Allowable maximum pressure 44 MPa 70 MPa 105 MPa 130 MPa
Flow rate settings range 0.0001-5.0000 mL/min
(1.0-44 MPa)
5.0001-10.0000 mL/min
(1.0-22 MPa)
0.0001-3.0000 mL/min
(1.0-70 MPa)
3.0001-5.0000 mL/min
(1.0-44 MPa)
5.0001-10.0000 mL/min
(1.0-22 MPa)
0.0001-3.0000 mL/min
(1.0-105 MPa)
3.0001-5.0000 mL/min
(1.0-80 MPa)
5.0001-10.0000 mL/min
(1.0-22 MPa)
0.0001-3.0000 mL/min
(1.0-130 MPa)
3.0001-5.0000 mL/min
(1.0-80 MPa)
5.0001-10.0000 mL/min
(1.0-22 MPa)
Flow rate accuracy ≤± 1% or ± 2 μL/min, whichever greater
(under specified conditions)
≤± 1% (under specified conditions)
Flow rate accuracy ≤ 0.06% RSD or 0.02 minSD, whichever greater
Gradient mode
 
High-pressure gradient (2 or 3 solvents) Quaternary low-pressure gradient High-pressure gradient (2 solvents (LC-40B XR standard) or 3 solvents) Quaternary low-pressure gradient (Only available for LC-40D XR) High-pressure gradient (2 or 3 solvents) Quaternary low-pressure gradient High-pressure gradient (2 solvents (LC-40B X3 standard) or 3 solvents) Quaternary low-pressure gradient (Only available for LC-40D X3)
Gradient range of set concentrations 0 to 100% (0.1% step)
Gradient range of set concentrations ± 0.5% (under specified conditions)
Wetted materials SUS316L, Hastelloy® C, PEEK, PTFE, Sapphire, Ruby SUS316L, Hastelloy C, PEEK, PE, Sapphire, Ruby
Available pH range 1 to 14
Automatic rinsing kit Option Standard equipment
Degassing unit 1 unit connectable LC-40D XR:1 unit connectable LC-40B XR: 2 units connectable 1 unit connectable
LC-40D X3:
1 unit connectable
LC-40B X3:
pre-installed (5 port builtin), 1 unit connectable
Dimensions  W 260×D 500×H 140 mm LC-40D X3:
W 260×D 500×H 140 mm
LC-40B X3:
W 260×D 500×H 210 mm
Weight 10 kg LC-40D XR:10 kg
LC-40B XR:13 kg
12 kg LC-40D X3:12 kg
LC-40B X3: 21 kg
Operating temperature range 4 to 35°C
Power supply AC 100-240 V、50/60 Hz
150 VA LC-40D XR:150VA
LC-40B XR:180 VA
150 VA LC-40D X3:150 VA
LC-40B X3:180 VA

For Research Use Only. Not for use in diagnostic procedures.