Nexera series

Autosampler

SIL-40C XR

 

  SIL-40
SIL-40C
SIL-40 XR
SIL-40C XR
SIL-40C XS SIL-40C X3
Injection method
 
Total-volume Injection (standard), loop injection (optional)
Allowable
maximum pressure
44 MPa 80 MPa 105 MPa 130 MPa
Injection volume 0.1-100 μL 0.1-50 μL
0.1 to 2000 μL (optional)
Injection volume
accuracy
≤ ± 1% (5 μL injection, n = 20)
Linearity ≥ 0.9999
Injection cycle time ≤ 6.7 seconds (under specified conditions)
Samples for processing   288( microtiter plate, 96 well × 3 plates)
  1152( microtiter plate, 384 well × 3 plates)
252( 1 mL sample vial, 84 × 3 plates)
  162( 1.5 mL sample vial, 54 × 3 plates)
84( 4 mL sample vial, 28 × 3 plates)
  36( 10 mL sample vial, 12 × 3 plates)
     72( 1.5 mL micro tube, 24 × 3 plates)) 
Injection volume reproducibility
 
RSD≦1.0% (0.5-0.9 μL)
RSD≦0.5% (1.0-1.9 μL)
 RSD≦0.25%( 2.0-4.9 μL)
           RSD≦0.15%( More than 5.0 μL)
       RSD<0.5% (typically, 0.5 μL)
         RSD<0.25%( typically, 1.0 μL)
Carryover
 
≤ 0.0025% (without rinse)
≤0.0005%
(with rinse, typically)
(under specified conditions)
≤ 0.0015%(without rinse)
≤ 0.0003%(with rinse, typically))
(under specified conditions)
Dip rinsing outside the needle
and injection port rinsing
 
Standard equipment
Pumping rinse
outside the needle
Option Standard equipment
Internal rinsing (3 dil) Option Standard equipment
Sample cooler
 
SIL-40:None
SIL-40C:Standard equipment
(Air-circulation temperature
control type)
 
SIL-40 XR:None
SIL-40C XR:Standard equipment
(Air-circulation temperature
control type)
 
Standard equipment
(Air-circulation temperature control type)
Sample cooler
temperature setting range
 
4 to 45°C
(Room temperature needs to be less than 30°C and humidity needs to be less than 70% to set 4°C)
Sample cooler
temperature accuracy
 
±2 ℃(sensor position ± 0.5°C)
Wetted material SUS316L, DLC, PEEK, GFP, PTFE, FEP, ETFE, sapphire, ceramics, PPS, FFKM
Available pH range 1 to 14
Dimensions  W 260×D 500×H 280 mm(SIL-40C/40C XR/40C XS/40C X3Protrusion adds 140 mm to the depth)
Weight SIL-40:17 kg
SIL-40C:24 kg
 
SIL-40 XR:17 kg
SIL-40C XR:24 kg
24 kg
Operating
temperature range
4 to 35 ℃
Power supply Cooler model AC 100-240 V, 50/60 Hz
Non cooler model SIL-40:150 VA
SIL-40C:400 VA
SIL-40 XR:150 VA
SIL-40C XR:400 VA
 

For Research Use Only. Not for use in diagnostic procedures.

This page may contain references to products that are not available in your country. Please contact us to check the availability of these products in your country.

Top of This Page